Installation Reports

TABLE OF CONTENTS
• Executive Summary
Introduction
The purpose of this work is to install the Vibration Isolation MB 3000 system for the Thermo Fisher Scientific Glacios 2 at Pfizer. Thermo Fisher Scientific hired VEC to install the Daeil MB-3000 vibration isolation platform and provide a report of the completed work.
Summary Results
Below is a summary of the results:
- MB3000 ON PASSED the Vibration specifications
• Introduction
The purpose of this work is to install the Vibration Isolation MB 3000 system for the Thermo
Fisher Scientific Glacios 2 at Pfizer. Thermo Fisher Scientific hired VEC to install the Daeil MB-
3000 vibration isolation platform and provide a report of the completed work.
• Product Details
| Part Delivered | Serial Number | Software |
|---|---|---|
| MB3000 | 230518R1 | Site-Specific Tune |
• Summary of Work
We installed the system by
- Positioning the Daeil under the microscope
We tested the system by
- Creating and uploading a site specific tune
- Collecting vibration data to verify the performance
We left the system
- In position and on
• Procedures
Vibration
VEC conducted vibration measurements in six directions. Acceleration measurements were taken at a bandwidth of 200 with units of um/s (RMS). The acceleration spectra with 200 bandwidths were recorded.
• Setup
| Vibration | |
|---|---|
| Bandwidth | 200 |
| Lines | 1600 |
| Window | Hanning |
| Coupling | ICP 2mA |
| Average Type | Stable |
| Duration | 60 |
| Engineering Units | um/s (RMS) |
• Specifications
VEC used the following specifications for the project.
Specification Summary
| Tool | Specification |
|---|---|
| Thermo Fisher Scientific Glacios 2 |
Vibration VC-F X & Y 1.56 um/s 1Hz - 100Hz VC-E Z 3.12 um/s 1Hz - 100Hz |
Detailed Specifications
Figure 1: Thermo Fisher Scientific Glacios 2 – 1 Vibration Specification
| System | Vertical | Left to right | Front to back |
|---|---|---|---|
| Glacios; TEM info Limit 0.23 nm | E | F | F |
| Glacios; STEM res. 0.23 nm | E | F | F |
• Results
Summary Results
Below is a summary of the results:
- MB3000 PASSED the Vibration specifications
Detailed Results
| Type | Position | Direction | Specification | Measurement | Result |
|---|---|---|---|---|---|
| Vibration | Floor | X-Axis | VC-F | VC-G | PASS |
| Y-Axis | VC-G | PASS | |||
| Z-Axis | VC-E | VC-D | PASS | ||
| Isolators | X-Axis | VC-F | VC-G | PASS | |
| Y-Axis | VC-G | PASS | |||
| Z-Axis | VC-E | VC-G | PASS |
• Data
Thermo Fisher Scientific Glacios 2
Vibration
Figure 2: Ground X-Axis
Figure 3: Isolators X Axis
Figure 4: Transmissibility X Axis
Figure 5: Ground Y-Axis
Figure 6: Isolators Y Axis
Figure 7: Transmissibility Y Axis
Figure 8: Ground Z-Axis
Figure 9: Isolators Z Axis
Figure 10: Transmissibility Z Axis
Figure 11: Final Results – SN