JEOL Korea Ltd. ETRI JEOL JBX-8100FS DVIA-P4000 Installation Report
Installation report for the DVIA-P4000 vibration isolation table, serial 260731R1, installed for a JEOL JBX-8100FS E-Beam Lithography system in the Lab Building 4 clean room at ETRI in Daejeon. After re-tuning with the customer equipment turned on (2026-11-10), the top-plate vibration met the JBX-8100FS allowable velocity curve over 1 ~ 100 Hz in all three axes, with a maximum of 5.6 % of the allowable limit (Vertical, 59.3 Hz).

ZEISS EVO 18 SEM DVIA-UB700 Installation Report
Re-tuning report of the DVIA-UB700 (S/N 190903R1) installed in 2019 at Shanghai Maritime Safety Administration, Nenjiang Road Base, supporting a ZEISS EVO 18 SEM. Vibration was measured in the Idle state in Autospectrum format, and the isolated platform passed the ZEISS EVO series specification in all three axes.

ZEISS Sigma 360 DVIA-ML1000 Installation Report
Installation report of DVIA-ML1000 (S/N 260424R6) for (End User: Fulian Technology (Shanxi) Co., Ltd.), serving a ZEISS Sigma 360 FE-SEM in China. After a 2nd tuning, the DVIA-ML1000 satisfied the vibration specification in all directions, reducing floor vibration that exceeded the specification in the Vertical direction.

ZEISS Gemini SEM 460 DVIA-ML1000 Installation Report
Installation report of DVIA-ML1000 (S/N 260529R6) for ZEISS (End User: Tier-1 Semiconductor Onyang), serving a ZEISS Gemini SEM 460 FE-SEM. After tuning, top-plate vibration on all X/Y/Z axes met the ZEISS Gemini SEM installation requirements, with maximums of 0.85 µm/s rms (Z), 1.85 µm/s rms (X) and 1.83 µm/s rms (Y).

Hitachi High-Tech SU-3800SEPlus DVIA-ML1000 Installation Report
DVIA-ML1000 installation report (serial 260728R1-1) for a Hitachi High-Tech SEM at Fujian Xiangfenghua New Energy Materials in Shanghai. After the 1st tuning trial on 2026.11.02, with the equipment in the IDLE state, both the floor and the DVIA-ML1000 passed the vibration specification on all three axes.

RAITH Voyager DVIA-ML1000 Installation Report
Installation report of DVIA-ML1000 (S/N 260824R1) for (End User: Institute of Semiconductors, Chinese Academy of Sciences, Beijing), serving a RAITH Voyager Electron Beam Lithography system. After tuning, the DVIA-ML1000 satisfied the vibration specification in all directions, reducing the floor's Vertical exceedance to within specification.