Skip to main content
Installation Report
11-09-2026

YIMO TECH ZEISS EVO 18 SEM DVIA-UB700 Installation Report

DVIA-ULF Series
Installation Report
YIMO TECH
Shanghai Maritime Safety Administration, Nenjiang Road Base
ZEISS
EVO 18
DVIA-UB700
190903R1

Overview

The DVIA-UB700 installed in Shanghai in 2019 (1st tuning: 2019.11.07) was re-tuned, and vibration measurement was performed.

Vibration measurement was conducted with the equipment in the Idle state.

Data is represented in Autospectrum format and reference material on the vibration level is provided.

Vibration Isolation System Information

Model: DVIA-UB700

Serial Number: 190903R1

Engineer

Younguk KIM · DAEIL SYSTEMS

Tuning Date

2026.11.09

End User

Shanghai Maritime Safety Administration, Nenjiang Road Base

Number of Tuning Trial

2nd

Location

Shanghai, China

Equipment

Manufacturer: ZEISS

Equipment: SEM

Model: EVO 18

Vibration Specification

Applied specification: ZEISS EVO series floor vibration specification (manufacturer) — less than 6 µm/s RMS up to 30 Hz and less than 12 µm/s RMS above 30 Hz, in three orthogonal axes.

DirectionFrequency [Hz]Allowable [µm/s RMS]
Horizontal (X, Y)1 – 306
30 – 8012
Vertical (Z)1 – 306
30 – 8012

Equipment Condition

Idle

Engineering Tool & Setup

ItemUI (built-in sensors)
Instrument · SoftwareDVIA-UB700 controller · DAEIL SYSTEMS Tuning Software (UI)
SensorBuilt-in velocity sensors (Top / Bottom)
Measurement PointTop = DVIA-UB700 isolated platform / Bottom = floor side (base) — measured simultaneously
AxisVertical / Left to Right / Front to Back — one axis at a time

Conclusion

Following the 2nd tuning of the DVIA-UB700 active vibration isolation system, the vibration on the isolated platform satisfies the ZEISS EVO series floor vibration specification in all three axes.

Measurement Summary

AxisResultStatus
FloorDVIA-UB700FloorDVIA-UB700
Vertical1.7 µm/s @ 3.0 Hz0.70 µm/s @ 3.0 Hz✓ PASS✓ PASS
Left to Right1.3 µm/s @ 4.7 Hz0.84 µm/s @ 2.4 Hz✓ PASS✓ PASS
Front to Back2.6 µm/s @ 3.2 Hz1.0 µm/s @ 2.9 Hz✓ PASS✓ PASS

Results are the peak values of the velocity Autospectrum (µm/s RMS). The floor and the isolated platform are each compared with the allowable level in Section 9.

Measurement Data

Vertical

Vertical – Autospectrum (SPEC: ZEISS EVO series)

Left to Right

Left to Right – Autospectrum (SPEC: ZEISS EVO series)

Front to Back

Front to Back – Autospectrum (SPEC: ZEISS EVO series)

Installation Photo

ZEISS EVO 18 on the DVIA-UB700

Reference

The table below classifies vibration levels based on IEST-RP-CC012.2.

Criterion CurveDescriptionAmplitude¹
(µm/s RMS)
Detail Size²μm
Workshop (ISO)Distinctly perceptible vibration. Appropriate to workshops and non-sensitive areas.800N/A
Office (ISO)Perceptible vibration. Appropriate to offices and non-sensitive areas.400N/A
Residential Area (ISO)Barely perceptible vibration. Appropriate to sleep areas in most instances. Usually adequate for computer equipment, hospital recovery rooms, semiconductor probe test equipment, and microscopes less than 40x.20075
Operating Theatre (ISO)Vibration not perceptible. Suitable in most instances for surgical suites, microscopes to 100x and for other equipment of low sensitivity.10025
VC-AAdequate in most instances for optical microscopes to 400x, microbalances, optical balances, proximity and projection aligners, etc.508
VC-BAppropriate for inspection and lithography equipment (including steppers) to 3μm line widths.253
VC-CAppropriate standard for optical microscopes to 1000x, lithography and inspection equipment (including moderately sensitive electron microscopes) to 1μm detail size. TFT-LCD stepper/scanner processes.12.51-3
VC-DSuitable in most instances for demanding equipment, including many electron microscopes (SEMs and TEMs) and E-Beam systems.6.250.1-0.3
VC-EA challenging criterion to achieve. Assumed to be adequate for the most demanding of sensitive systems including long path, laser-based, small target systems, E-Beam lithography systems working at nanometer scales, and other systems requiring extraordinary dynamic stability.3.12<0.1
VC-FAppropriate for extremely quiet research spaces; generally difficult to achieve in most instances, especially cleanrooms. Not recommended for use as a design criterion, only for evaluation.1.56N/A
VC-GAppropriate for extremely quiet research spaces; generally difficult to achieve in most instances, especially cleanrooms. Not recommended for use as a design criterion, only for evaluation.0.78N/A

¹ As measured in one-third octave bands of frequency over the frequency range 8 to 80 Hz (VC-A and VC-B) or 1 to 80 Hz (VC-C through VC-G).

² The detail size refers to line width in the case of microelectronics fabrication, the particle (cell) size in the case of medical and pharmaceutical research, etc. It is not relevant to imaging associated with probe technologies, AFMs, and nanotechnology.

The information given in this table is for guidance only. In most instances, it is recommended that the advice of someone knowledgeable about applications and vibration requirements of the equipment and processes be sought.

Share this Case Study

Case Study Information

Category
Installation Report
Date11-09-2026
Tags
DVIA-ULF Series
Installation Report
YIMO TECH
Shanghai Maritime Safety Administration, Nenjiang Road Base
ZEISS
EVO 18
DVIA-UB700
190903R1