Hanyang University ZEISS Sigma 300 DVIA-UB700 Installation Report
Contents
Prepared for
Hanyang University(Seoul)
Prepared by
Engineer: Seounghoon Cheon
Tuning date: 22.12.15
Report written date: 22.12.19
Date of business trip : 17.12.27~17.12.29
Overview
The DVIA-UB700 active vibration isolation platform is appropriately installed and functioning as intended.
System Information
Model: DVIA-UB700(910x1140x139H)
Serial Number: 221024R4
Engineer
Seounghoon Cheon, from DAEIL SYSTEMS
Tuning Date
December 15, 2022
Location
Hanyang university, Seoul
Equipment
ZEISS SEM Sigma 300
Measurement Device
7.1) Data Physics
Hardware: Quattro, Serial Number: 22436
Software: SignalCalc ACE
7.2) Measurement Setting
-Bandwidth: 0 – 80 Hz
-Lines: 800
-Window: Hanning
-Averaging: FFT Spectrum Averaging
-Engineering Units: m/s^2
7.3) Sensor
PCB Accelerometer
Model: 393B05
Summary of Vibration Results
| Measurement Point | Z-axis (Vertical)<br>1–10 Hz | Z-axis (Vertical)<br>12.5–80 Hz | X-axis (Left to Right)<br>1–10 Hz | X-axis (Left to Right)<br>12.5–80 Hz | Y-axis (Front to Back)<br>1–10 Hz | Y-axis (Front to Back)<br>10–80 Hz |
|---|---|---|---|---|---|---|
| Floor | F | B | F | E | G | E |
| Active | G | E | G | G | G | G |
Data and Image (Equipment Turn off)
Floor Z-axis (Vertical) VC curves
Top Z-axis (Vertical) VC curves
Z-axis (Vertical) Transmissibility
Floor X-axis (Left to Right) VC curves
Top X-axis (Left to Right) VC curves
X-axis (Left to Right) Transmissibility
Floor Y-axis (Front to Back) VC curves
Top Y-axis (Front to Back) VC curves
Y-axis (Front to Back) Transmissibility
Reference
Notes:
1. VC-A/B is measured in 1/3 octave bands from 8-80 Hz, VC-C through VC-G from 1-80 Hz.
2. Detail size refers to line widths in microelectronics manufacturing or particle sizes in medical research.