Following the installation of the DVIA-ML1000 active vibration isolation system for the Raith Voyager Electron Beam Lithography system at the Institute of Semiconductors, Chinese Academy of Sciences (Beijing), tuning and vibration measurement were performed on 2026.11.02.
Vibration was measured on the floor and on the DVIA-ML1000 in three axes (Vertical, Left to Right, Front to Back), with the equipment in the Idle state.
The results are presented in Autospectrum format and evaluated against the equipment vibration specification (Section 9). Reference material on vibration levels is provided in Section 16.
Vibration Isolation System Information
Model: DVIA-ML1000
Serial Number: 260824R1
Engineer
Taehun Kim · DAEIL SYSTEMS
Tuning Date
2026.11.02
End User
Institute of Semiconductors, Chinese Academy of Sciences
Number of Tuning Trial
1st
Site
Beijing
Equipment
Manufacturer: RAITH
Equipment: Electron Beam Lithography system
Model: Voyager
Vibration Specification
| Axis | Frequency [Hz] | Allowable [µm/s RMS] |
|---|---|---|
| Vertical & Horizontal | 1 ~ 16 | 0.8 |
| 16 ~ | 1 |
Equipment Condition
Idle
Tuning Request
Conclusion
The DVIA-ML1000 installed for the Raith Voyager Electron Beam Lithography system at the Institute of Semiconductors, Chinese Academy of Sciences (Beijing) was tuned on 2026.11.02, with the equipment in the Idle state.
Vibration was measured on the floor and on the DVIA-ML1000 in the Vertical, Left to Right and Front to Back directions. The DVIA-ML1000 satisfies the vibration specification in all directions. The floor exceeded the specification in the Vertical direction, and that vibration was reduced to within the specification on the DVIA-ML1000.
Measurement Summary
| Measurement Point | Vibration Specifications | Axis | Result | |
|---|---|---|---|---|
| Floor | DVIA-ML1000 | |||
| 1. Floor 2. DVIA-ML1000 | See Section 9 | Vertical | ✗ FAIL | ✓ PASS |
| Left to Right | ✓ PASS | ✓ PASS | ||
| Front to Back | ✓ PASS | ✓ PASS | ||
Measurement Data Obtained via UI Program
Vertical — Autospectrum
Left to Right — Autospectrum
Front to Back — Autospectrum
Installation Photo
Equipment
Reference
The table below classifies vibration levels based on IEST-RP-CC012.2.
Notes:
1. As measured in one-third octave bands of frequency over the frequency 8 to 80 Hz (VC-A and VC-B) or 1 to 80 Hz (VC-C through VC-G).
2. The detail size refers to the width in the case of microelectronics fabrication, the particle (cell) size in the case of medical and pharmaceutical research, etc. It is not relevant to imaging associated with probe technologies, AFMs, and nanotechnology.
The information given in this table is for guidance only. In most instances, it is recommended that the advice of someone knowledgeable about applications and vibration requirements of the equipment and processes be sought.