Installation Report
11-02-2026

YIMOTECH ZEISS Sigma 360 DVIA-ML1000 Installation Report

DVIA-M Series
Installation Report
YIMOTECH
Fulian Technology (Shanxi) Co., Ltd.
ZEISS
Sigma 360
DVIA-ML1000
260424R6

Overview

Following the 1st tuning of the DVIA-ML1000 in Fulian Technology (Shanxi) Co., Ltd., the customer reported that the system was wavering and not stable, while the vibration isolation performance was good. At the customer’s request, a 2nd tuning and vibration measurement were performed on 2026.11.05.

Vibration measurement was conducted with the equipment in the Turned off state.

Data is represented in Autospectrum format and reference material on the vibration level is provided.

Vibration Isolation System Information

Model: DVIA-ML1000

Serial Number: 260424R6

Engineer

Taehun Kim · DAEIL SYSTEMS

Tuning Date

2026.11.02

End User

Fulian Technology (Shanxi) Co., Ltd. (富联科技)

Number of Tuning Trial

2nd

Site

China

Equipment

Manufacturer: ZEISS

Equipment: FE-SEM

Model: Sigma 360

Vibration Specification

Horizontal

Frequency [Hz]Allowable [mm/s² RMS]
1 ~ 100.04
10 ~ 200.17
20 ~700.3
70 ~2

Vertical

Frequency [Hz]Allowable [mm/s² RMS]
1 ~ 80.03
8 ~ 450.15
45 ~2

Equipment Condition

Turned off

Tuning Request

Conclusion

The customer reported that the DVIA-ML1000 installed for the ZEISS Sigma 360 FE-SEM in China was wavering and not stable. A 2nd tuning was performed on 2026.11.05, with the equipment in the Turned off state.

Vibration was measured on the floor and on the DVIA-ML1000 in the Vertical, Left to Right and Front to Back directions. The DVIA-ML1000 satisfies the vibration specification in all directions. The floor exceeded the specification in the Vertical direction, and that vibration was reduced to within the specification on the DVIA-ML1000.

Measurement Summary

Measurement PointVibration SpecificationsAxisResult
FloorDVIA-ML1000
1. Floor
2. DVIA-ML1000
See Section 9Vertical✗ FAIL✓ PASS
Left to Right✓ PASS✓ PASS
Front to Back✓ PASS✓ PASS

Measurement Data Obtained via UI Program

Vertical — Autospectrum

Left to Right — Autospectrum

Front to Back — Autospectrum

Installation Photo

Equipment

Reference

The table below classifies vibration levels based on IEST-RP-CC012.2.

Criterion CurveDescriptionAmplitude¹μm/s
(μin/s)
Detail Size²μm
Workshop (ISO)Distinctly perceptible vibration. Appropriate to workshops and non-sensitive areas.800 (32,000)N/A
Office (ISO)Perceptible vibration. Appropriate to offices and non-sensitive areas.400 (16,000)N/A
Residential Area (ISO)Barely perceptible vibration. Appropriate to sleep areas in most instances. Usually adequate for computer equipment, hospital recovery rooms, semiconductor probe test equipment, and microscopes less than 40x.200 (8,000)75
Operating Theatre (ISO)Vibration not perceptible. Suitable in most instances for surgical suites, microscopes to 100x and for other equipment of low sensitivity.100 (4,000)25
VC-AAdequate in most instances for optical microscopes to 400x, microbalances, optical balances, proximity and projection aligners, etc.50 (2,000)8
VC-BAppropriate for inspection and lithography equipment (including steppers) to 3μm line widths.25 (1,000)3
VC-CAppropriate standard for optical microscopes to 1000x, lithography and inspection equipment (including moderately sensitive electron microscopes) to 1μm detail size. TFT-LCD stepper/scanner processes.12.5 (500)1-3
VC-DSuitable in most instances for demanding equipment, including many electron microscopes (SEMs and TEMs) and E-Beam systems.6.25 (250)0.1-0.3
VC-EA challenging criterion to achieve. Assumed to be adequate for the most demanding of sensitive systems including long path, laser-based, small target systems, E-Beam lithography systems working at nanometer scales, and other systems requiring extraordinary dynamic stability.3.12 (125)<0.1
VC-FAppropriate for extremely quiet research spaces; generally difficult to achieve in most instances, especially cleanrooms. Not recommended for use as a design criterion, only for evaluation.1.56 (62.5)N/A
VC-GAppropriate for extremely quiet research spaces; generally difficult to achieve in most instances, especially cleanrooms. Not recommended for use as a design criterion, only for evaluation.0.78 (31.3)N/A

Notes:

1. As measured in one-third octave bands of frequency over the frequency 8 to 80 Hz (VC-A and VC-B) or 1 to 80 Hz (VC-C through VC-G).

2. The detail size refers to the width in the case of microelectronics fabrication, the particle (cell) size in the case of medical and pharmaceutical research, etc. It is not relevant to imaging associated with probe technologies, AFMs, and nanotechnology.

The information given in this table is for guidance only. In most instances, it is recommended that the advice of someone knowledgeable about applications and vibration requirements of the equipment and processes be sought.