YIMOTECH ZEISS Sigma 360 DVIA-ML1000 Installation Report
Contents
Overview
Following the 1st tuning of the DVIA-ML1000 in Fulian Technology (Shanxi) Co., Ltd., the customer reported that the system was wavering and not stable, while the vibration isolation performance was good. At the customer’s request, a 2nd tuning and vibration measurement were performed on 2026.11.05.
Vibration measurement was conducted with the equipment in the Turned off state.
Data is represented in Autospectrum format and reference material on the vibration level is provided.
Vibration Isolation System Information
Model: DVIA-ML1000
Serial Number: 260424R6
Engineer
Taehun Kim · DAEIL SYSTEMS
Tuning Date
2026.11.02
End User
Fulian Technology (Shanxi) Co., Ltd. (富联科技)
Number of Tuning Trial
2nd
Site
China
Equipment
Manufacturer: ZEISS
Equipment: FE-SEM
Model: Sigma 360
Vibration Specification
Horizontal
| Frequency [Hz] | Allowable [mm/s² RMS] |
|---|---|
| 1 ~ 10 | 0.04 |
| 10 ~ 20 | 0.17 |
| 20 ~70 | 0.3 |
| 70 ~ | 2 |
Vertical
| Frequency [Hz] | Allowable [mm/s² RMS] |
|---|---|
| 1 ~ 8 | 0.03 |
| 8 ~ 45 | 0.15 |
| 45 ~ | 2 |
Equipment Condition
Turned off
Tuning Request
Conclusion
The customer reported that the DVIA-ML1000 installed for the ZEISS Sigma 360 FE-SEM in China was wavering and not stable. A 2nd tuning was performed on 2026.11.05, with the equipment in the Turned off state.
Vibration was measured on the floor and on the DVIA-ML1000 in the Vertical, Left to Right and Front to Back directions. The DVIA-ML1000 satisfies the vibration specification in all directions. The floor exceeded the specification in the Vertical direction, and that vibration was reduced to within the specification on the DVIA-ML1000.
Measurement Summary
| Measurement Point | Vibration Specifications | Axis | Result | |
|---|---|---|---|---|
| Floor | DVIA-ML1000 | |||
| 1. Floor 2. DVIA-ML1000 | See Section 9 | Vertical | ✗ FAIL | ✓ PASS |
| Left to Right | ✓ PASS | ✓ PASS | ||
| Front to Back | ✓ PASS | ✓ PASS | ||
Measurement Data Obtained via UI Program
Vertical — Autospectrum
Left to Right — Autospectrum
Front to Back — Autospectrum
Installation Photo
Equipment
Reference
The table below classifies vibration levels based on IEST-RP-CC012.2.
Notes:
1. As measured in one-third octave bands of frequency over the frequency 8 to 80 Hz (VC-A and VC-B) or 1 to 80 Hz (VC-C through VC-G).
2. The detail size refers to the width in the case of microelectronics fabrication, the particle (cell) size in the case of medical and pharmaceutical research, etc. It is not relevant to imaging associated with probe technologies, AFMs, and nanotechnology.
The information given in this table is for guidance only. In most instances, it is recommended that the advice of someone knowledgeable about applications and vibration requirements of the equipment and processes be sought.