Vibration Isolation Case Studies
DAEIL SYSTEMS engineers active vibration isolation systems for the world's most demanding nanoscale imaging and metrology environments.
Series
Equipment Type
Manufacturer
All Case Studies

ZEISS FE-SEM Sigma 360 DVIA-ULF700 Installation Report
DVIA-ULF700 installation report (serial 260605R7) for a ZEISS FE-SEM Sigma 360 at Hubei Jinquan in Hubei. A 2nd tuning was carried out on 2026.09.21 at the customer's request after the initial tuning on 2026.09.08. With the equipment in the Idle state, vibration levels on the DVIA-ULF700 met the specification on all three axes, while the floor vibration exceeded the specification on all axes.

ZEISS EVO 10 SEM DVIA-ULF700 Installation Report
DVIA-ULF700 installation report (serial 260827R1) for a ZEISS EVO 10 SEM at End User Hebei Hengke Vibration Damping Equipment Co., Ltd. in Hebei. Tuning was carried out on 2026.09.14 and vibration was measured with the equipment in the IDLE state at two points, the installation floor and the top surface of the DVIA-ULF700. All three axes (Vertical, Left to Right, Front to Back) passed the specified vibration criterion at both measurement points, the floor and the DVIA-ULF700. Published figures: the tuning request work order, the per-axis Autospectrum traces captured from the UI program, and one installation photo.
Tier-1 Semiconductor Cheonan ZEISS X-ray Microscope DVIA-ULF1500 Installation Report
Tier-1 Semiconductor Cheonan (C3 line, 6th floor mass production quality team cleanroom) ZEISS Xradia 730 Versa X-ray microscope DVIA-ULF1500 installation report. After relocation and installation, tuning and vibration measurement were performed with the equipment powered on and active isolation running. At the central point between the two ULF modules the floor vibration was reduced within the VC-C grade on all axes. Per-axis VC Curves are published.

ZEISS FE-SEM Sigma360 DVIA-ULF700 Installation Report
DVIA-ULF700 installation report (serial 260528R9) for a ZEISS FE-SEM Sigma360 at End User School of Life and Health Sciences, Hainan Normal University in Haikou. Tuning was carried out on 2026.09.11 and vibration was measured with the equipment in the IDLE state at two points, the installation floor and the top surface of the DVIA-ULF700. All three axes (Vertical, Left to Right, Front to Back) passed the specified vibration criterion at both measurement points, the floor and the DVIA-ULF700.

ASE Korea TESCAN FIB-SEM SOLARIS X2 GMH DVIA-ML1000 Installation Report
DVIA-ML1000 installation report (serial 260622R4) for a TESCAN FIB-SEM SOLARIS X2 GMH installed at ASE Korea PB1, 1CMD QO Lab, Electron Microscopy Room. Tuning was carried out on 2026.09.11 and vibration was measured with the equipment in the IDLE state. All three axes met the reference specification and achieved a PASS rating, with the vertical axis improving from VC-B to VC-E, the left-to-right axis from VC-E to VC-G, and the front-to-back axis from VC-B to VC-F.

Pharrics BV MPI Corporation Probe Station MPI TS150 DVIA-T67 (260402R1) Installation Report
Installation report for the DVIA-T67 (serial 260402R1) installed for Pharrics BV in Leuven under an MPI Corporation MPI TS150 probe station. With no vibration specification defined, tuning and vibration measurements were performed with the probe station mounted on the system in the idle state. The results are presented as per-axis Autospectrum and VC Curve plots together with the IEST-RP-CC012.2 vibration criteria reference.