Skip to main content
Installation Report
11-02-2026

Hitachi High-Tech SU-3800SEPlus DVIA-ML1000 Installation Report

DVIA-M Series
Installation Report
YIMOTECH
Fujian Xiangfenghua New Energy Materials
Hitachi High-Tech
SEM
SU-3800SEPlus
DVIA-ML1000
260728R1-1
Shanghai

Overview

Following the installation of the DVIA-ML1000 active vibration isolation system for the Hitachi High-Tech SU8600 SEM at Fujian Xiangfenghua New Energy Materials Co., Ltd. (Shanghai), tuning and vibration measurement were performed on 2026.11.02.

Vibration was measured on the floor and on the DVIA-ML1000 in three axes (Vertical, Left to Right, Front to Back), with the equipment in the IDLE state.

The results are evaluated against the equipment vibration specification (Section 9). Reference material on vibration levels is provided in Section 16.

Vibration Isolation System Information

Model: DVIA-ML1000

Serial Number: 260728R1-1

Engineer

Taehun Kim · DAEIL SYSTEMS

Tuning Date

2026.11.02

End User

Fujian Xiangfenghua New Energy Materials

Number of Tuning Trial

1st

Site

Shanghai

Equipment

Manufacturer: Hitachi High-Tech

Equipment: SEM

Model: SU-3800SEPlus

Vibration Specification

Horizontal

Frequency [Hz]Allowable [µm pk-pk]
21.5
31.6
53.0
82.0
101.3

Vertical

Frequency [Hz]Allowable [µm pk-pk]
210.0
35.0
52.0
81.5
103.0

Equipment Condition

IDLE

Tuning Request

Conclusion

The DVIA-ML1000 installed for the Hitachi High-Tech SU8600 SEM at Fujian Xiangfenghua New Energy Materials Co., Ltd. (Shanghai) was tuned on 2026.11.02, with the equipment in the IDLE state.

Vibration was measured on the floor and on the DVIA-ML1000 in the Vertical, Left to Right and Front to Back directions. The DVIA-ML1000 satisfies the vibration specification in all directions.

Measurement Summary

Measurement PointVibration SpecificationsAxisResult
FloorDVIA-ML1000
1. Floor
2. DVIA-ML1000
See Section 9Vertical✓ PASS✓ PASS
Left to Right✓ PASS✓ PASS
Front to Back✓ PASS✓ PASS

Measurement Data Obtained via UI Program

Vertical

Vertical — Autospectrum

Left to Right

Left to Right — Autospectrum

Front to Back

Front to Back — Autospectrum

Installation Photo

Equipment

Reference

The table below classifies vibration levels based on IEST-RP-CC012.2.

Criterion CurveDescriptionAmplitude¹μm/s
(μin/s)
Detail Size²μm
Workshop (ISO)Distinctly perceptible vibration. Appropriate to workshops and non-sensitive areas.800 (32,000)N/A
Office (ISO)Perceptible vibration. Appropriate to offices and non-sensitive areas.400 (16,000)N/A
Residential Area (ISO)Barely perceptible vibration. Appropriate to sleep areas in most instances. Usually adequate for computer equipment, hospital recovery rooms, semiconductor probe test equipment, and microscopes less than 40x.200 (8,000)75
Operating Theatre (ISO)Vibration not perceptible. Suitable in most instances for surgical suites, microscopes to 100x and for other equipment of low sensitivity.100 (4,000)25
VC-AAdequate in most instances for optical microscopes to 400x, microbalances, optical balances, proximity and projection aligners, etc.50 (2,000)8
VC-BAppropriate for inspection and lithography equipment (including steppers) to 3μm line widths.25 (1,000)3
VC-CAppropriate standard for optical microscopes to 1000x, lithography and inspection equipment (including moderately sensitive electron microscopes) to 1μm detail size. TFT-LCD stepper/scanner processes.12.5 (500)1-3
VC-DSuitable in most instances for demanding equipment, including many electron microscopes (SEMs and TEMs) and E-Beam systems.6.25 (250)0.1-0.3
VC-EA challenging criterion to achieve. Assumed to be adequate for the most demanding of sensitive systems including long path, laser-based, small target systems, E-Beam lithography systems working at nanometer scales, and other systems requiring extraordinary dynamic stability.3.12 (125)<0.1
VC-FAppropriate for extremely quiet research spaces; generally difficult to achieve in most instances, especially cleanrooms. Not recommended for use as a design criterion, only for evaluation.1.56 (62.5)N/A
VC-GAppropriate for extremely quiet research spaces; generally difficult to achieve in most instances, especially cleanrooms. Not recommended for use as a design criterion, only for evaluation.0.78 (31.3)N/A

¹ As measured in one-third octave bands of frequency over the frequency range 8 to 80 Hz (VC-A and VC-B) or 1 to 80 Hz (VC-C through VC-G).

² The detail size refers to line width in the case of microelectronics fabrication, the particle (cell) size in the case of medical and pharmaceutical research, etc. It is not relevant to imaging associated with probe technologies, AFMs, and nanotechnology.

The information given in this table is for guidance only. In most instances, it is recommended that the advice of someone knowledgeable about applications and vibration requirements of the equipment and processes be sought.

Share this Case Study

Case Study Information

Category
Installation Report
Date11-02-2026
Tags
DVIA-M Series
Installation Report
YIMOTECH
Fujian Xiangfenghua New Energy Materials
Hitachi High-Tech
SEM
SU-3800SEPlus
DVIA-ML1000
260728R1-1
Shanghai