Hitachi High-Tech SU-3800SEPlus DVIA-ML1000 Installation Report
Contents
Overview
Following the installation of the DVIA-ML1000 active vibration isolation system for the Hitachi High-Tech SU8600 SEM at Fujian Xiangfenghua New Energy Materials Co., Ltd. (Shanghai), tuning and vibration measurement were performed on 2026.11.02.
Vibration was measured on the floor and on the DVIA-ML1000 in three axes (Vertical, Left to Right, Front to Back), with the equipment in the IDLE state.
The results are evaluated against the equipment vibration specification (Section 9). Reference material on vibration levels is provided in Section 16.
Vibration Isolation System Information
Model: DVIA-ML1000
Serial Number: 260728R1-1
Engineer
Taehun Kim · DAEIL SYSTEMS
Tuning Date
2026.11.02
End User
Fujian Xiangfenghua New Energy Materials
Number of Tuning Trial
1st
Site
Shanghai
Equipment
Manufacturer: Hitachi High-Tech
Equipment: SEM
Model: SU-3800SEPlus
Vibration Specification
Horizontal
| Frequency [Hz] | Allowable [µm pk-pk] |
|---|---|
| 2 | 1.5 |
| 3 | 1.6 |
| 5 | 3.0 |
| 8 | 2.0 |
| 10 | 1.3 |
Vertical
| Frequency [Hz] | Allowable [µm pk-pk] |
|---|---|
| 2 | 10.0 |
| 3 | 5.0 |
| 5 | 2.0 |
| 8 | 1.5 |
| 10 | 3.0 |
Equipment Condition
IDLE
Tuning Request
Conclusion
The DVIA-ML1000 installed for the Hitachi High-Tech SU8600 SEM at Fujian Xiangfenghua New Energy Materials Co., Ltd. (Shanghai) was tuned on 2026.11.02, with the equipment in the IDLE state.
Vibration was measured on the floor and on the DVIA-ML1000 in the Vertical, Left to Right and Front to Back directions. The DVIA-ML1000 satisfies the vibration specification in all directions.
Measurement Summary
| Measurement Point | Vibration Specifications | Axis | Result | |
|---|---|---|---|---|
| Floor | DVIA-ML1000 | |||
| 1. Floor 2. DVIA-ML1000 | See Section 9 | Vertical | ✓ PASS | ✓ PASS |
| Left to Right | ✓ PASS | ✓ PASS | ||
| Front to Back | ✓ PASS | ✓ PASS | ||
Measurement Data Obtained via UI Program
Vertical
Vertical — Autospectrum
Left to Right
Left to Right — Autospectrum
Front to Back
Front to Back — Autospectrum
Installation Photo
Equipment
Reference
The table below classifies vibration levels based on IEST-RP-CC012.2.
¹ As measured in one-third octave bands of frequency over the frequency range 8 to 80 Hz (VC-A and VC-B) or 1 to 80 Hz (VC-C through VC-G).
² The detail size refers to line width in the case of microelectronics fabrication, the particle (cell) size in the case of medical and pharmaceutical research, etc. It is not relevant to imaging associated with probe technologies, AFMs, and nanotechnology.
The information given in this table is for guidance only. In most instances, it is recommended that the advice of someone knowledgeable about applications and vibration requirements of the equipment and processes be sought.
Share this Case Study
Case Study Information
Related Case Studies

Fujian University of Technology Hitachi SU7000 SEM DVIA-ML1000 Installation Report

Changdi New Materials Technology Hitachi High-Tech SU3800SE FE-SEM DVIA-ML1000 Installation Report