Skip to main content

Vibration Isolation Case Studies

DAEIL SYSTEMS engineers active vibration isolation systems for the world's most demanding nanoscale imaging and metrology environments.

Series

Equipment Type

Manufacturer

Filtered Case Studies

441 case studies
ZEISS Gemini 300 DVIA-MB1000 (201123R3) Internal Program Performance — Shenzhen — 2021.04.28

Field handout — DVIA-MB1000 serial 201123R3 (P210) for ZEISS Gemini 300 in Shenzhen; end user Tencent. Earlier transmissibility summary (21.03.09, Hubei, ZEISS Evo 18) and MB1000 internal-program Z/X/Y plots from the Word package. Ornamental JPEG after the setup title line is not republished.

Novartis Thermo Fisher Scientific Glacios DVIA-MB3000 (201012R5) Installation Report — 2021.04.21

DVIA-MB3000 (serial 201012R5, internal P15) remote VEC worksheet for Thermo Fisher Scientific Glacios at Novartis (MA): transmissibility and VC-class summary tables, MB1000 internal program axis plots, and VEC measurement graphics. Setup and measurement 2021-04-21; reporting 2021-04-22; operator Doohyun Kim (DAEIL SYSTEMS, research engineer). Historical business-trip window 2017-12-27~2017-12-29 per report header.

Thermo Fisher Scientific NXQ8000 Mask Aligner DVIA-MB1000 remote VEC (210302R3) Installation Report — 2021.04.20

Remote VEC worksheet for DVIA-MB1000 (serial 210302R3, internal P224) under THERMO FISCHER NXQ8000 Mask Aligner at Thermo Fisher Scientific, CA (B1). Operator Doohyun Kim; measurement 2021-04-08; setup and reporting dates as stated on the form; business trip window per header. Publishes equipment photograph, MB1000 internal program axis plots, and VEC measurement axis plots extracted from Word. The source handout has no Reference / Generic Vibration Criteria section — none was added.

LB Semicon TESCAN MIRA DVIA-MB1000 (MB1000 P220) Installation Report — 2021.04.16

MB1000 internal-program worksheet (P220) for TESCAN MIRA at LB Semicon (2F annex laboratory): VC-Class summary table, B&K measurement-chain figures, MB1000 internal-program Z/X/Y transmissibility plots, and floor versus isolated-top VC comparison spectra by axis. Generic Vibration Criteria table and standalone VC curve chart embedded in Word are omitted on the public page in favor of the standard markdown Reference table.

Shagang Group ZEISS Crossbeam 550 DVIA-MB1000 (201028R1 · P202, remote tuning) — 2021.04.12

DVIA-MB1000 serial 201028R1 (project P202) for ZEISS Crossbeam 550 reported by /channel partner Shagang Group (Zhangjiagang installation line). Remote performance worksheet: dates 2021-04-12 (setup and measurement) / 2021-04-13 (reporting), operator Doohyun Kim (DAEIL SYSTEMS research engineer). Publishes the setup photograph from under the worksheet title, transmissibility summary at 2 / 5 / 10 Hz, and MB1000 internal-program captures under Z / X / Y axis captions as in Word.

TPK Holding Co., Ltd Thermo Fisher Helios G5 DVIA-MB1000 (201214R4 · P216) Remote measurement — Internal program plots — 2021.04.09

DVIA-MB1000 serial 201214R4 (project P216) for Thermo Fisher Helios G5 reported by /channel partner; installation-site table cites Xiamen TPK Holding 4F with measurement dated 2021.04.08 also at Xiamen. Figures follow Z axis, X axis, and Y axis as in Word; the photo embedded directly under (P216) Set up is ornamental and is not republished.