Vibration Isolation Case Studies

DAEIL SYSTEMS engineers active vibration isolation systems for the world's most demanding nanoscale imaging and metrology environments.

Series

Equipment Type

Manufacturer

All Case Studies

1246 case studies
J&L Tech DVIA-ULF1500 Installation Report — 250904R1
Installation Report
DVIA-ULF Series
12-22-2025

J&L Tech DVIA-ULF1500 Installation Report — 250904R1

DVIA-ULF1500 (serial 250904R1) at J&L Tech Ansan Branch R&D lab: PASS/floor-vs-table VC summary, nine VC / autospectrum / transmissibility plots, and generic VC criterion table reference.

YIMO ZEISS Sigma360 DVIA-ULF700 Installation Report
Installation Report
DVIA-ULF Series
12-18-2025

YIMO ZEISS Sigma360 DVIA-ULF700 Installation Report

YIMO ZEISS Sigma 360 DVIA-ULF700 Installation Report

Kunshan DeKaiYiTe ZEISS FE-SEM Sigma 360 DVIA-ULF700 Installation Report

DVIA-ULF700 (serial 250923R3) 1st tuning with ZEISS FE-SEM Sigma 360 for Kunshan DeKaiYiTe (Location N/A in the report). Publishes vibration specification capture, VC / autospectrum / transmissibility plots for three axes, equipment photo, and VC reference table (no decorative header or EMF curve upload). Report written January 7, 2026.

Safran JEOL JSM-IT800 DVIA-ML1000 Installation Report
Installation Report
DVIA-M Series
12-17-2025

Safran JEOL JSM-IT800 DVIA-ML1000 Installation Report

Safran Transmission System LAB, Colombes — JEOL JSM-IT800 on DVIA-ML1000 (250829R2).

Li Auto ZEISS Sigma 360 DVIA-ULF700 Installation Report
Installation Report
DVIA-ULF Series
12-12-2025

Li Auto ZEISS Sigma 360 DVIA-ULF700 Installation Report

Li Auto (Shanghai) ZEISS FE-SEM Sigma 360 DVIA-ULF700 installation and vibration tuning report.

YIMO Beijing University of Aeronautics and Astronautics Unisoku STM DVIA-ML1000 (250730R1-1) Installation Report

DVIA-ML1000 installation report — Beijing University of Aeronautics and Astronautics Unisoku STM (serial 250730R1-1). Overview documents Z-axis VCM issues (Dec 3 / Dec 11 visit), interim tuning with Z2 off and Z3/Z4 matrix sign reversal, and acceptable overall performance with slight horizontal-axis degradation pending full VCM repair. Transmissibility plots (vertical, left-to-right, front-to-back), tuning-request capture, and equipment photo.