Vibration Isolation Case Studies
DAEIL SYSTEMS engineers active vibration isolation systems for the world's most demanding nanoscale imaging and metrology environments.
Series
Equipment Type
Manufacturer
Filtered Case Studies

Thermo Fisher Scientific TFS Helios (Brigham Young University) DVIA-ML1000 (250708R4-3) Installation Report
Vibration isolation installation report for DVIA-ML1000 serial 250708R4-3 beneath a Thermo Fisher Scientific Helios platform at BYU Electron Microscopy Facility: VC-F vertical and VC-G horizontal vibration criteria; floor versus DVIA comparison plots with transmissibility; onsite installation and serial photos.

DVIA-ML1000 (250227R2) Installation Report
DVIA-ML1000 installation report (serial 250227R2).
Tier-1 Semiconductor DSR ZEISS Versa 630 X-ray Microscope DVIA-ML3000 (250625R2) Installation Report
DVIA-ML3000 (250625R2) installation report — Tier-1 Semiconductor DSR A Tower 5F analysis lab, ZEISS X-ray Microscope Versa 630. Equipment vibration specification as published UI capture; VC / Autospectrum / Transmissibility (merged strip pairs where required); Generic Vibration Criteria as markdown table.

ZEISS FE-SEM Sigma 560 DVIA-ML1000 (250611R3) Installation Report
DVIA-ML1000 installation report — ZEISS FE-SEM Sigma 560 (serial 250611R3). Overview states the platform is appropriately installed and functioning as intended. Prepared for, End User, and Location are N/A on the form. Equipment condition: The equipment is installed/IDLE; Tuning Request N/A. UI-program transmissibility (vertical, left-to-right, front-to-back), on-site equipment photo; Reference uses the standard Generic Vibration Criteria markdown table. No vibration specification section and no measurement summary table on the published form.
Tier-1 Semiconductor ZEISS Versa 620 X-ray Microscope DVIA-MB3000 (170515R4) Installation Report — 2025.07.08
DVIA-MB3000 (serial 170515R4) on-site tuning and vibration worksheet dated 2025.07.08 at Tier-1 Semiconductor Cheonan campus (equipment Turned on/IDLE): VC curves, autospectra, and transmissibility for floor vs DVIA-MB3000 platform, measurement summary table (vertical floor FAIL / platform PASS at reported frequencies), equipment vibration specification excerpt (VC-C), and Reference as a markdown Generic Vibration Criteria table (no separate reference figure). Engineer Seo Jong-hwa (DAEIL SYSTEMS); report written 2025.07.10.

Supergravity Laboratory QIYUE SS-HV800 FIB-SEM DVIA-M3000 (240612R3) Installation Report
DVIA-M3000 (240612R3) installation report — Supergravity Laboratory, Hangzhou, China; QIYUE TECHNOLOGY FIB-SEM SS-HV800. Published Transmissibility plots (Vertical, Left to Right, Front to Back), Tuning Request capture, and equipment photograph. Generic vibration criteria presented as markdown table.