Vibration Isolation Case Studies
DAEIL SYSTEMS engineers active vibration isolation systems for the world's most demanding nanoscale imaging and metrology environments.
Series
Equipment Type
Manufacturer
Filtered Case Studies

SpaceX Thermo Fisher Scientific FIB-SEM DVIA-U1000 Installation Report
SpaceX Thermo Fisher Scientific FIB-SEM DVIA-U1000 Installation Report

Shenzhen BYD EVO 10 DVIA-U1000 (241120R3) Installation Report — Shenzhen
DVIA-U1000 compact installation worksheet (serial 241120R3) for Shenzhen BYD with ZEISS SEM EVO 10 in Shenzhen. Setup and reporting 25.03.18; business-trip line 17.12.27~17.12.29; engineers Chaewon Lee and Jonghwa Suh. Overview measurement dated 2025.03.18 (Thursday). Publishes setup photograph and Vertical / Left to Right / Front to Back internal-program performance screenshots.

Shenzhen BYD ZEISS Sigma 360 DVIA-U1000 (240308R3) Installation Report — Shenzhen
DVIA-U1000 compact installation worksheet (serial 240308R3) at Shenzhen BYD with ZEISS FE-SEM Sigma 360 in Shenzhen. Setup and reporting 25.02.20; operators Chaewon Lee and Jonghwa Suh from DAEIL SYSTEMS; business-trip line 17.12.27~17.12.29. Overview measurement dated 2025.02.20 (Thursday). Published installation photograph plus U1000 internal-program Vertical / Left to Right / Front to Back screenshots in Word segment order.

Korea Institute of Science and Technology (L7-222) Olympus BX51 DVIA-UD350 (220127R1-2) Installation Report
DVIA-UD350 (serial 220127R1-2) vibration inspection/tuning report at Korea Institute of Science and Technology Building L7, Room 222 for an Olympus BX51 microscope. Measurements taken with equipment turned off; VC curves, autospectra, and transmissibility for vertical/left‑to‑right/front‑to‑back; Dataphysics QUATTRO/SignalCalc ACE. Y1 sensor replacement noted. Decorative cover bitmap omitted; VC reference reproduced as markdown only.

Tier-1 Semiconductor DVIA-U850 (230724R1) service report — Cheonan facility
Follow-up DVIA-U850 vibration service report (serial from source filename 230724R1) documenting October 2024 measurements at Tier-1 Semiconductor Cheonan-area line: fan/cable remediation, VC-curve plotting, settling-time runs (Align/PBI/BLT), transmissibility, and comparative prior Suwon relocation context. Decorative cover/header artwork omitted; generic VC criterion handout under Reference is reproduced as markdown only.

Shanghai Public Security Bureau HITACHI SU7000 DVIA-U1000 (230330R1) Installation Report — Shanghai
DVIA-U1000 compact installation worksheet (serial 230330R1) at Shanghai Public Security Bureau with HITACHI SU7000 in Shanghai. Setup 24.08.05, reporting 24.08.07; Overview measurement dated 2024.08.05 (Monday); operator Chaewon Lee from DAEIL SYSTEMS; business-trip line 17.12.27~17.12.29. Publication includes installation photograph plus U1000 internal-program Vertical / Left to Right / Front to Back screenshots in Word segment order.