Skip to main content
Installation Report
DVIA-M Series
07-04-2016

Dongjin Semichem Incheon Hitachi SU8010 DVIA-MB1000 (MB1000 P21) vibration environment measurement — 2016-07-04

DVIA-M Series
Installation Report
DVIA-MB1000
MB1000
P21
Hitachi
SU8010
SEM
Dongjin Semichem
Incheon

MB1000 P21 vibration environment measurement result (Dongjin Semichem)

Equipment installation photograph

Field visit date: 2016.07.01 (Fri)

Report date: 2016.07.04 (Mon)

Field staff: Kim Du-hyeon (Deputy Manager), Lim Jong-wook (Assistant Manager), DAEIL SYSTEMS

Overview

After installing the vibration isolation platform, the vibration environment was measured to confirm there are no issues for equipment use.

Measurement date: 2016. 07. 01 (Fri), Measured by: Lim Jong-wook (Assistant Manager), DAEIL SYSTEMS

Measurement site: Dongjin Semichem Incheon factory

Equipment model: HITACHI SEM ( SU8010 )

Measurement conditions and results

Measurement condition: The MB1000 top surface and the building floor were measured.

Summary of measurement results

Test ConditionVC-Class ZVC-Class XVC-Class Y
FloorBCC
Above isolation platform topFFF

Overall performance is favorable in all directions.

VC-Class graphs and built-in program transmissibility are attached.

Vibration environment VC-CLASS graphs

Z axis vibration level

X axis vibration level

Y axis vibration level

Built-in program transmissibility

Z axis transmissibility

X axis transmissibility

Y axis transmissibility

Share this Case Study

Case Study Information

Category
Installation Report
SeriesDVIA-M Series
Date07-04-2016
Tags
DVIA-M Series
Installation Report
DVIA-MB1000
MB1000
P21
Hitachi
SU8010
SEM
Dongjin Semichem
Incheon