Vibration Isolation Case Studies

DAEIL SYSTEMS engineers active vibration isolation systems for the world's most demanding nanoscale imaging and metrology environments.

Series

Equipment Type

Manufacturer

Filtered Case Studies

184 case studies
Institute for Basic Science (IBS) OLYMPUS_U-CMAD3 UD350 P8 Performance Report (April 2019)

Performance worksheet dated April 18, 2019 (April 17, 2019 site visit): UD350 vibration isolation supporting OLYMPUS_U-CMAD3 at the Institute for Basic Science 2nd-floor analysis laboratory (P8). VC-Class summary table (floor C/F/F vs isolator top G/G/G) under stated conditions, U350 internal-program Z/X/Y screenshots, and octave-band plots (Floor vs isolator top). Decorative title-block raster from the DOCX is omitted from publication.

Xiamen University of Technology Zeiss Sigma 500 DVIA-UB700 (P9) China trip installation report — 2019.01.07

DVIA-UB700 worksheet P9 from the 2019-01-07 Chinese trip report for Zeiss Sigma 500 at Xiamen University of Technology (SEM Lab, 1F). Trip date 19.01.03; trip personnel Choi, Sung Won (최성원). Installation result summary VC-Class table and follow-up notes follow the form verbatim; MB1000 internal program and vibration-environment UI captures are published under Z, X, and Y with §2a embed mapping (image3–8). Decorative image1 and the title-block raster under the centred report title (image2) are omitted from the public page.

Shanghai Institute of Technical Physics DVIA-UB700 (180927R1) Installation Report

DVIA-UB700 (S/N 180927R1) installation worksheet at Shanghai Institute of Technical Physics (SITP); 1st floor site; operator/manager Sung Won Choi from DAEIL SYSTEMS — setup date 18.12.19; report dated 18.12.26 per worksheet. Publication includes MB1000 internal-program Z/X/Y performance plots and octave-band (Floor vs Above Active) Z/X/Y plots; title-block images (media/image1, image2) are omitted.

Little Swan UB1000 P6 China Business Trip Report — Zeiss EVO 18 SEM (2018-05-10)

UB1000_P6 setup business trip report for Zeiss EVO 18 SEM at Little Swan. Trip date 2018.05.10; report written 2018.05.14; engineer CHOI SUNG WON. VC-Class summary table (ACTIVE Base Z C / Top E; X and Y E/E), installation results narrative, internal-program Z/X/Y ACTIVE transmissibility plots, and floor vibration-environment Z/X/Y 1/3-octave plots. Decorative header PNG omitted.

Tier-1 Semiconductor Bruker VERTEX 80 DVIA-U1000 (UO1000_P5) Installation Report — Seoul

DVIA-U1000 compact worksheet UO1000_P5 for Bruker VERTEX 80 at Tier-1 Semiconductor (Seoul Samsung Hospital B4 animal laboratory). Business trip and report written 18.02.06; operator Hyungmoon Choi. VC-Class summary table (building floor vs isolator top plate) and follow-up notes per the form; MB1000 internal-program and octave-band (floor vs isolator top) Z/X/Y screenshots per §2a embed mapping. Decorative header bitmap (image1) omitted.

Saeron Technology UB1000 P3 demo result — in-house SEM vibration survey (2016-10-21)

Demo report dated 21 October 2016 for UB1000 P3 at Saeron Technology in Uiwang, Gyeonggi-do, comparing on-site vibration and 100,000× SEM imaging against the existing MB1000 platform on an in-house SEM. Travelers CHOI HYOUNGMUN and LIM JONG UK; measurer LIM JONG UK. Publication includes the in-house SEM photograph, floor versus platform VC-class Z/X/Y plots, and ACTIVE image-comparison captures for MB1000 and UB1000. Decorative header PNG omitted.