Installation Report
10-20-2026

ZEISS Sigma 360 FE-SEM DVIA-ULF700 Installation Report

DVIA-ULF Series
Installation Report
Xintaijia Optoelectronics
ZEISS
Sigma 360
DVIA-ULF700
260902R5

Overview

Following the installation of DVIA-ULF700 at Linhai, Zhejiang, tuning and vibration measurement were performed.

Vibration measurement was conducted with the equipment in the Idle state.

Data is represented in Autospectrum format and reference material on the vibration level is provided.

Vibration Isolation System Information

Model: DVIA-ULF700

Serial Number: 260902R5

Engineer

Taehun KIM · DAEIL SYSTEMS

Tuning Date

2026.10.20

End User

Xintaijia Optoelectronics

Number of Tuning Trial

1st

Site

Linhai, Zhejiang

Equipment

Manufacturer: ZEISS

Equipment: FE-SEM

Model: Sigma 360

Vibration Specification

Horizontal

Frequency [Hz]Allowable [mm/s² RMS]
10.04
100.04
100.17
200.17
200.3
700.3
702
1002

Vertical

Frequency [Hz]Allowable [mm/s² RMS]
10.03
80.03
80.15
450.15
452
1002

Equipment Condition

Idle

Tuning Request

Conclusion

The DVIA-ULF700 (S/N 260902R5) was installed and tuned at the Linhai, Zhejiang site.

A structural resonance near 10 Hz (horizontal, X-axis) from the passive section was suppressed by controller tuning. Attenuation at this frequency did not reach -20 dB; the result was accepted.

In Idle state, the floor exceeded the referenced specification (Section 9) on all axes, while the DVIA-ULF700 met it on all axes.

Note: Measured before the SEM covers were installed. A re-check is recommended after final assembly.

Measurement Summary

Measurement PointVibration SpecificationsAxisResult
FloorDVIA-ULF700
1. Floor
2. DVIA-ULF700
See Section 9Vertical✗ FAIL✓ PASS
Left to Right✗ FAIL✓ PASS
Front to Back✗ FAIL✓ PASS

Measurement Data Obtained via UI Program

Vertical

Vertical — Autospectrum

Left to Right

Left to Right — Autospectrum

Front to Back

Front to Back — Autospectrum

Installation Photo

Reference

The table below classifies vibration levels based on IEST-RP-CC012.2.

Criterion CurveDescriptionAmplitude¹μm/s
(μin/s)
Detail Size²μm
Workshop (ISO)Distinctly perceptible vibration. Appropriate to workshops and non-sensitive areas.800 (32,000)N/A
Office (ISO)Perceptible vibration. Appropriate to offices and non-sensitive areas.400 (16,000)N/A
Residential Area (ISO)Barely perceptible vibration. Appropriate to sleep areas in most instances. Usually adequate for computer equipment, hospital recovery rooms, semiconductor probe test equipment, and microscopes less than 40x.200 (8,000)75
Operating Theatre (ISO)Vibration not perceptible. Suitable in most instances for surgical suites, microscopes to 100x and for other equipment of low sensitivity.100 (4,000)25
VC-AAdequate in most instances for optical microscopes to 400x, microbalances, optical balances, proximity and projection aligners, etc.50 (2,000)8
VC-BAppropriate for inspection and lithography equipment (including steppers) to 3μm line widths.25 (1,000)3
VC-CAppropriate standard for optical microscopes to 1000x, lithography and inspection equipment (including moderately sensitive electron microscopes) to 1μm detail size. TFT-LCD stepper/scanner processes.12.5 (500)1-3
VC-DSuitable in most instances for demanding equipment, including many electron microscopes (SEMs and TEMs) and E-Beam systems.6.25 (250)0.1-0.3
VC-EA challenging criterion to achieve. Assumed to be adequate for the most demanding of sensitive systems including long path, laser-based, small target systems, E-Beam lithography systems working at nanometer scales, and other systems requiring extraordinary dynamic stability.3.12 (125)<0.1
VC-FAppropriate for extremely quiet research spaces; generally difficult to achieve in most instances, especially cleanrooms. Not recommended for use as a design criterion, only for evaluation.1.56 (62.5)N/A
VC-GAppropriate for extremely quiet research spaces; generally difficult to achieve in most instances, especially cleanrooms. Not recommended for use as a design criterion, only for evaluation.0.78 (31.3)N/A

Notes:

1. As measured in one-third octave bands of frequency over the frequency 8 to 80 Hz (VC-A and VC-B) or 1 to 80 Hz (VC-C through VC-G).

2. The detail size refers to the width in the case of microelectronics fabrication, the particle (cell) size in the case of medical and pharmaceutical research, etc. It is not relevant to imaging associated with probe technologies, AFMs, and nanotechnology.

The information given in this table is for guidance only. In most instances, it is recommended that the advice of someone knowledgeable about applications and vibration requirements of the equipment and processes be sought.