Qingdao Daeil Shanghai Huayuyi Marble Table + Interferometer DVIA-ML3000S (251229R1-1) Installation Report

DVIA-ML Series
Installation Report
Qingdao Daeil
Shanghai Huayuyi
Marble Table
Interferometer
DVIA-ML3000S
251229R1-1
Zhejiang
Shanghai
China

Overview

Remote pre-tuning (预调试) of DVIA-ML3000S (S/N 251229R1-1) was performed at the pre-tuning site in Zhejiang, before the system is relocated to the final installation site in Shanghai.

Performance verification data was obtained via the UI program. Data is represented in Autospectrum and VC Curve format, and reference material on the vibration level is provided.

Vibration Isolation System Information

Model: DVIA-ML3000S (External Controller)

Serial Number: 251229R1-1

Engineer

Gyumin.park · DAEIL SYSTEMS

Tuning Date

2026.10.16

End User

Shanghai Huayuyi (上海华雨益)

Number of Tuning Trial

1st (Pre-tuning, Remote)

Site

Pre-tuning: Zhejiang, China / Final installation: Shanghai, China

Equipment

Manufacturer: -

Equipment: Marble Table + Interferometer (大理石+干涉仪)

Model: -

Vibration Specification

N/A (No vibration specification defined)

Acceptance criterion: No oscillation (抖动) of the system with Active ON.

Equipment Condition

Pre-tuning condition: interferometer not installed, marble load approx. 2,800 kg (almost the same as the final marble weight).

Tuning Request

Conclusion

Remote pre-tuning and vibration measurement of DVIA-ML3000S (S/N 251229R1-1) were completed.

No oscillation was observed with Active ON during the pre-tuning.

Floor vibration at the pre-tuning site was VC-C (Vertical), VC-E (Left to Right) and VC-E (Front to Back), and the isolator top reached VC-G level in all three axes.

Average attenuation (1–80 Hz): Vertical -27 dB / Left to Right -29 dB / Front to Back -19 dB.

The final performance may differ due to reassembly, the installation of the interferometer and side boxes, and the floor condition of the final site.

Measurement Summary

Measurement PointVibration SpecificationsAxisResult
FloorDVIA-ML3000S
1. Floor
2. DVIA-ML3000S
N/AVerticalVC-CVC-G
Left to RightVC-EVC-G
Front to BackVC-EVC-G

※ No vibration specification defined; results are shown as VC grade (1–80 Hz, 1/3 octave).

Measurement Data Obtained via UI Program

Vertical

Vertical — Autospectrum

Vertical — VC Curves

Left to Right

Left to Right — Autospectrum

Left to Right — VC Curves

Front to Back

Front to Back — Autospectrum

Front to Back — VC Curves

Installation Photo

Installation (Pre-tuning site, Zhejiang)

Nameplate (DVIA-ML3000S, S/N 251229R1-1)

Reference

The table below classifies vibration levels based on IEST-RP-CC012.2.

Criterion CurveDescriptionAmplitude
µm/s (µin/s)
Detail Size
µm
Workshop (ISO)Distinctly perceptible vibration. Appropriate to workshops and non-sensitive areas.800 (32,000)N/A
Office (ISO)Perceptible vibration. Appropriate to offices and non-sensitive areas.400 (16,000)N/A
Residential Area (ISO)Barely perceptible vibration. Appropriate to sleep areas in most instances. Usually adequate for computer equipment, hospital recovery rooms, semiconductor probe test equipment, and microscopes less than 40x.200 (8,000)75
Operating Theatre (ISO)Vibration not perceptible. Suitable in most instances for surgical suites, microscopes to 100x and for other equipment of low sensitivity.100 (4,000)25
VC-AAdequate in most instances for optical microscopes to 400x, microbalances, optical balances, proximity and projection aligners, etc.50 (2,000)8
VC-BAppropriate for inspection and lithography equipment (including steppers) to 3μm line widths.25 (1,000)3
VC-CAppropriate standard for optical microscopes to 1000x, lithography and inspection equipment (including moderately sensitive electron microscopes) to 1μm detail size. TFT-LCD stepper/scanner processes.12.5 (500)1-3
VC-DSuitable in most instances for demanding equipment, including many electron microscopes (SEMs and TEMs) and E-Beam systems.6.25 (250)0.1-0.3
VC-EA challenging criterion to achieve. Assumed to be adequate for the most demanding of sensitive systems including long path, laser-based, small target systems, E-Beam lithography systems working at nanometer scales, and other systems requiring extraordinary dynamic stability.3.12 (125)<0.1
VC-FAppropriate for extremely quiet research spaces; generally difficult to achieve in most instances, especially cleanrooms. Not recommended for use as a design criterion, only for evaluation.1.56 (62.5)N/A
VC-GAppropriate for extremely quiet research spaces; generally difficult to achieve in most instances, especially cleanrooms. Not recommended for use as a design criterion, only for evaluation.0.78 (31.3)N/A

Notes:

- As measured in one-third octave bands of frequency over the frequency 8 to 80 Hz (VC-A and VC-B) or 1 to 80 Hz (VC-C through VC-G).

- The detail size refers to the width in the case of microelectronics fabrication, the particle (cell) size in the case of medical and pharmaceutical research, etc. It is not relevant to imaging associated with probe technologies, AFMs, and nanotechnology.

The information given in this table is for guidance only. In most instances, it is recommended that the advice of someone knowledgeable about applications and vibration requirements of the equipment and processes be sought.