Thermo Fisher Scientific Apreo SEM DVIA-ML1000 Installation Report
Contents
Overview
Following the installation of DVIA-ML1000 at Zhenjiang(镇江), tuning and vibration measurement were performed. Vibration measurement was conducted with the equipment in the Idle state.
Data is represented in VC Curve format and reference material on the vibration level is provided.
Vibration Isolation System Information
Model: DVIA-ML1000
Serial Number: 251218R1-6
Engineer
Younguk KIM · DAEIL SYSTEMS
Tuning Date
2026.08.11
End User
jiangsu tiangong precision tools co ltd(江苏天工精密工具有限公司)
Number of Tuning Trial
1st
Site
Zhenjiang(镇江)
Equipment
Manufacturer: Thermo Fisher Scientific
Equipment: SEM
Model: Apreo
Vibration Specification
Horizontal
| Frequency [Hz] | Allowable [µm/s RMS] |
|---|---|
| 1 | 1.5625 |
| 100 | 1.5625 |
Vertical
| Frequency [Hz] | Allowable [µm/s RMS] |
|---|---|
| 1 | 1.5625 |
| 100 | 1.5625 |
Equipment Condition
Idle
Tuning Request
Measurement Summary
| Measurement Point | Vibration Specifications | Axis | Result | |
|---|---|---|---|---|
| Floor | DVIA-ML1000 | |||
| 1. Floor 2. DVIA-ML1000 | See Section 9 | Vertical | ✗ FAIL | ✓ PASS |
| Left to Right | ✗ FAIL | ✓ PASS | ||
| Front to Back | ✗ FAIL | ✓ PASS | ||
Measurement Data Obtained via UI Program
Vertical
Vertical — VC Curves
Left to Right
Left to Right — VC Curves
Front to Back
Front to Back — VC Curves
Installation Photo
Equipment
Reference
The table below classifies vibration levels based on IEST-RP-CC012.2.
¹ As measured in one-third octave bands of frequency over the frequency 8 to 80 Hz (VC-A and VC-B) or 1 to 80 Hz (VC-C through VC-G).
² The detail size refers to the width in the case of microelectronics fabrication, the particle (cell) size in the case of medical and pharmaceutical research, etc. It is not relevant to imaging associated with probe technologies, AFMs, and nanotechnology.
The information given in this table is for guidance only. In most instances, it is recommended that the advice of someone knowledgeable about applications and vibration requirements of the equipment and processes be sought.