Skip to main content
Installation Report
DVIA-M Series
06-22-2026

YIMO TECH Fuyang University of Technology Thermo Fisher Apreo 2C SEM DVIA-ML1000 Installation Report

DVIA-M Series
Installation Report
YIMO TECH
Fuyang University of Technology
Thermo Fisher Scientific
Apreo 2C
SEM
DVIA-ML1000
251218R1-3
Fuyang

Overview

The DVIA-ML1000 active vibration isolation platform is appropriately installed and functioning as intended

Vibration Isolation System Information

Model: DVIA-ML1000

Serial Number: 251218R1-3

Engineer

Juhyeok.kim · DAEIL SYSTEMS

Installation Date

2026.06.22

Report Written Date

2026.06.22

End User

Fuyang University of Technology

Number of Tuning Trial

1st

Location

Fuyang in China

Equipment

Manufacturer: Thermo Fisher Scientific

Model: Apreo 2C

Tuning Request

Vibration Specification

Thermo Fisher Scientific conducts site surveys using SE-Tools to measure floor vibration, magnetic fields, and acoustic noise, and provides site-specific ambient vibration requirements through a site survey report. Therefore, the vibration specifications may vary depending on the survey results.

Measurement Summary

Measurement PointVibration SpecificationsAxisResult
FloorDVIA-ML1000
1. Floor
2. DVIA-ML1000
See Section9VerticalVC-GVC-G
Left to RightVC-GVC-G
Front to BackVC-GVC-G

Measurement Data

Vertical Autospectrum

Vertical VC Curves

Vertical Transmissibility

Left to Right Autospectrum

Left to Right VC Curves

Left to Right Transmissibility

Front to Back Autospectrum

Front to Back VC Curves

Front to Back Transmissibility

Equipment picture

Reference

The table below classifies vibration levels based on IEST-RP-CC012.2.

Criterion CurveDescription of UseAmplitude (μm/s RMS)Detail Size (μm)
Workshop (ISO)Distinctly perceptible vibration. Appropriate to workshops and non-sensitive areas.800N/A
Office (ISO)Perceptible vibration. Appropriate to offices and non-sensitive areas.400N/A
Residential Area (ISO)Barely perceptible vibration. Appropriate to sleep areas in most instances. Usually adequate for computer equipment, hospital recovery rooms, semiconductor probe test equipment, and microscopes less than 40x.20075
Operating Theatre (ISO)Vibration not perceptible. Suitable in most instances for surgical suites, microscopes to 100x and for other equipment of low sensitivity.10025
VC-AAdequate in most instances for optical microscopes to 400x, microbalances, optical balances, proximity and projection aligners, etc.508
VC-BAppropriate for inspection and lithography equipment (including steppers) to 3μm line widths.253
VC-CAppropriate standard for optical microscopes to 1000x, lithography and inspection equipment (including moderately sensitive electron microscopes) to 1μm detail size. TFT-LCD stepper/scanner processes.12.51–3
VC-DSuitable in most instances for demanding equipment, including many electron microscopes (SEMs and TEMs) and E-Beam systems.6.250.1–0.3
VC-EA challenging criterion to achieve. Assumed to be adequate for the most demanding of sensitive systems including long path, laser-based, small target systems. E-Beam lithography systems working at nanometer scales and other systems requiring extraordinary dynamic stability.3.12< 0.1
VC-FAppropriate for extremely quiet research spaces; generally difficult to achieve in most instances, especially cleanrooms. Not recommended for use as a design criterion, only for evaluation.1.56N/A
VC-GAppropriate for extremely quiet research spaces; generally difficult to achieve in most instances, especially cleanrooms. Not recommended for use as a design criterion, only for evaluation.0.78N/A

Notes:

1. As measured in one-third octave bands of frequency over the frequency 8 to 80 Hz (VC-A and VC-B) or 1 to 80 Hz (VC-C through VC-G).

2. The detail size refers to line width in the case of microelectronics fabrication, the particle (cell) size in the case of medical and pharmaceutical research, etc.

3. The detail size is not relevant to imaging associated with probe technologies, AFMs, and nanotechnology.