YIMO TECH Fuyang University of Technology Thermo Fisher Apreo 2C SEM DVIA-ML1000 Installation Report
Contents
Overview
The DVIA-ML1000 active vibration isolation platform is appropriately installed and functioning as intended
Vibration Isolation System Information
Model: DVIA-ML1000
Serial Number: 251218R1-3
Engineer
Juhyeok.kim · DAEIL SYSTEMS
Installation Date
2026.06.22
Report Written Date
2026.06.22
End User
Fuyang University of Technology
Number of Tuning Trial
1st
Location
Fuyang in China
Equipment
Manufacturer: Thermo Fisher Scientific
Model: Apreo 2C
Tuning Request
Vibration Specification
Thermo Fisher Scientific conducts site surveys using SE-Tools to measure floor vibration, magnetic fields, and acoustic noise, and provides site-specific ambient vibration requirements through a site survey report. Therefore, the vibration specifications may vary depending on the survey results.
Measurement Summary
| Measurement Point | Vibration Specifications | Axis | Result | |
|---|---|---|---|---|
| Floor | DVIA-ML1000 | |||
| 1. Floor 2. DVIA-ML1000 | See Section9 | Vertical | VC-G | VC-G |
| Left to Right | VC-G | VC-G | ||
| Front to Back | VC-G | VC-G | ||
Measurement Data
Vertical Autospectrum
Vertical VC Curves
Vertical Transmissibility
Left to Right Autospectrum
Left to Right VC Curves
Left to Right Transmissibility
Front to Back Autospectrum
Front to Back VC Curves
Front to Back Transmissibility
Equipment picture
Reference
The table below classifies vibration levels based on IEST-RP-CC012.2.
Notes:
1. As measured in one-third octave bands of frequency over the frequency 8 to 80 Hz (VC-A and VC-B) or 1 to 80 Hz (VC-C through VC-G).
2. The detail size refers to line width in the case of microelectronics fabrication, the particle (cell) size in the case of medical and pharmaceutical research, etc.
3. The detail size is not relevant to imaging associated with probe technologies, AFMs, and nanotechnology.