YIMO TECH Broadcast and Television Metrology Thermo Fisher Talos FIB-SEM DVIA-ML1000 Installation Report
Contents
Overview
The DVIA-ML1000 active vibration isolation platform is appropriately installed and functioning as intended.
Vibration Isolation System Information
Model: DVIA-ML1000
Serial Number: 251206R2-9
Engineer
Juhyeok.kim DAEIL SYSTEMS
Tuning Date
Jun 05, 2026
End User
Broadcast and Television Metrology
Number of Tuning Trial
1st
Location
Wuxi, China
Equipment
Manufacturer: Thermo Fisher Scientific
Equipment: FIB-SEM
Model: Talos
Vibration Specification
Thermo Fisher Scientific conducts site surveys using SE-Tools to measure floor vibration, magnetic fields, and acoustic noise, and provides site-specific ambient vibration requirements through a site survey report. Therefore, the vibration specifications may vary depending on the survey results.
Equipment Condition
The equipment is turned on/IDLE
Tuning Request
Measurement Summary
| Measurement Point | Status | Vibration Specification | Axis | Measurement Data(RMS, μm/s) | Measurement Result | ||
|---|---|---|---|---|---|---|---|
| Floor | DVIA-ML1000 | Floor | DVIA-ML1000 | ||||
| 1.Floor 2.DVIA-ML1000 | Equipment is Turned on/IDLE | See Section 9 | Vertical | 4.62 @ 15.5 Hz | 0.11 @ 15.5 Hz | ✓ PASS | ✓ PASS |
| Left to Right | 0.71 @ 4.5 Hz | 0.14 @ 4.5 Hz | ✓ PASS | ✓ PASS | |||
| Front to Back | 0.58 @ 3.5 Hz | 0.034 @ 3.5 Hz | ✓ PASS | ✓ PASS | |||
Measurement Data Obtained via UI Program
Vertical VC Curves
Vertical Autospectrum
Vertical Transmissibility
Left to Right VC curves
Left to Right Autospectrum
Left to Right Transmissibility
Front to Back VC Curves
Front to Back Autospectrum
Front to Back Transmissibility
Reference
Notes:
1. As measured in one-third octave bands of frequency over the frequency 8 to 80 Hz (VC-A and VC-B) or 1 to 80 Hz (VC-C through VC-G).
2. The detail size refers to line width in the case of microelectronics fabrication, the particle (cell) size in the case of medical and pharmaceutical research, etc.
3. The detail size is not relevant to imaging associated with probe technologies, AFMs, and nanotechnology.