Ningbo University of Technology Hitachi High-Tech FE-SEM SU8600 DVIA-ML1000 Installation Report
Contents
Overview
The DVIA-ML1000 active vibration isolation platform is appropriately installed and functioning as intended
Vibration Isolation System Information
Model: DVIA-ML1000
Serial Number: 250718R2
Engineer
Juhyeok.kim · DAEIL SYSTEMS
Installation Date
2026.06.05
End User
Ningbo University of Technology
Number of Tuning Trial
1st
Location
Ningbo · in China
Equipment
Manufacturer: Hitachi High-Tech
Equipment: FE-SEM
Model: SU8600
Tuning Request
Vibration Specification
Table 1-2-2 Allowable vibration
Allowable value (X, Y)
| Frequency [Hz] | Amplitude [μm p-p] |
|---|---|
| 1.0 | 8.0 or less |
| 1.5 | 3.0 or less |
| 2.0 | 1.4 or less |
| 2.4 | 1.4 or less |
| 3.0 | 3.0 or less |
| 5.0 | 4.0 or less |
| 6.0 | 3.5 or less |
| 10.0 | 3.5 or less |
Allowable value (Z)
| Frequency [Hz] | Amplitude [μm p-p] |
|---|---|
| 1.0 | 20 or less |
| 2.0 | 10 or less |
| 3.0 | 5.0 or less |
| 4.0 | 2.7 or less |
| 5.0 | 1.8 or less |
| 6.0 | 1.0 or less |
| 7.0 | 1.0 or less |
| 10.0 | 1.0 or less |
Measurement Summary
| Measurement Point | Vibration Specifications | Axis | Result | |
|---|---|---|---|---|
| Floor | DVIA-ML1000 | |||
| 1. Floor 2. DVIA-ML1000 | See Section 9 | Vertical | ✗ FAIL | ✓ PASS |
| Left to Right | ✗ FAIL | ✓ PASS | ||
| Front to Back | ✗ FAIL | ✓ PASS | ||
Measurement Data
Vertical
Vertical — Autospectrum
Vertical — VC Curves
Vertical — Transmissibility
Left to Right
Left to Right — Autospectrum
Left to Right — VC Curves
Left to Right — Transmissibility
Front to Back
Front to Back — Autospectrum
Front to Back — VC Curves
Front to Back — Transmissibility
Equipment picture
Reference
The table below classifies vibration levels based on IEST-RP-CC012.2.
1. As measured in one-third octave bands of frequency over the frequency 8 to 80 Hz (VC-A and VC-B) or 1 to 80 Hz (VC-C through VC-G).
2. The detail size refers to line width in the case of microelectronics fabrication, the particle (cell) size in the case of medical and pharmaceutical research, etc.
3. The detail size is not relevant to imaging associated with probe technologies, AFMs, and nanotechnology.