Installation Report
06-05-2026

Ningbo University of Engineering Hitachi High-Tech FE-SEM SU8600 DVIA-ML1000 Installation Report

DVIA-M Series
Installation Report
Ningbo University of Engineering
Hitachi High-Tech
FE-SEM
SU8600
DVIA-ML1000
250718R2
Ningbo

Overview

The DVIA-ML1000 active vibration isolation platform is appropriately installed and functioning as intended

Vibration Isolation System Information

Model: DVIA-ML1000

Serial Number: 250718R2

Engineer

Juhyeok.kim · DAEIL SYSTEMS

Installation Date

2026.06.05

End User

Ningbo University of Engineering

Number of Tuning Trial

1st

Location

Ningbo · in China

Equipment

Manufacturer: Hitachi High-Tech

Equipment: FE-SEM

Model: SU8600

Tuning Request

Vibration Specification

Table 1-2-2 Allowable vibration

Allowable value (X, Y)

Frequency [Hz]Amplitude [μm p-p]
1.08.0 or less
1.53.0 or less
2.01.4 or less
2.41.4 or less
3.03.0 or less
5.04.0 or less
6.03.5 or less
10.03.5 or less

Allowable value (Z)

Frequency [Hz]Amplitude [μm p-p]
1.020 or less
2.010 or less
3.05.0 or less
4.02.7 or less
5.01.8 or less
6.01.0 or less
7.01.0 or less
10.01.0 or less

Measurement Summary

Measurement PointVibration SpecificationsAxisResult
FloorDVIA-ML1000
1. Floor
2. DVIA-ML1000
See Section 9VerticalFAILPASS
Left to RightFAILPASS
Front to BackFAILPASS

Measurement Data

Vertical

Vertical — Autospectrum

Vertical — VC Curves

Vertical — Transmissibility

Left to Right

Left to Right — Autospectrum

Left to Right — VC Curves

Left to Right — Transmissibility

Front to Back

Front to Back — Autospectrum

Front to Back — VC Curves

Front to Back — Transmissibility

Equipment picture

Reference

The table below classifies vibration levels based on IEST-RP-CC012.2.

Criterion CurveDescription of UseAmplitude(μm/s RMS)Detail Size(μm)
Workshop (ISO)Distinctly perceptible vibration. Appropriate to workshops and non-sensitive areas.800N/A
Office (ISO)Perceptible vibration. Appropriate to offices and non-sensitive areas.400N/A
Residential Area (ISO)Barely perceptible vibration. Appropriate to sleep areas in most instances. Usually adequate for computer equipment, hospital recovery rooms, semiconductor probe test equipment, and microscopes less than 40x.20075
Operating Theatre (ISO)Vibration not perceptible. Suitable in most instances for surgical suites, microscopes to 100x and for other equipment of low sensitivity.10025
VC-AAdequate in most instances for optical microscopes to 400x, microbalances, optical balances, proximity and projection aligners, etc.508
VC-BAppropriate for inspection and lithography equipment (including steppers) to 3μm line widths.253
VC-CAppropriate standard for optical microscopes to 1000x, lithography and inspection equipment (including moderately sensitive electron microscopes) to 1μm detail size. TFT-LCD stepper/scanner processes.12.51–3
VC-DSuitable in most instances for demanding equipment, including many electron microscopes (SEMs and TEMs) and E-Beam systems.6.250.1–0.3
VC-EA challenging criterion to achieve. Assumed to be adequate for the most demanding of sensitive systems including long path, laser-based, small target systems. E-Beam lithography systems working at nanometer scales and other systems requiring extraordinary dynamic stability.3.12< 0.1
VC-FAppropriate for extremely quiet research spaces; generally difficult to achieve in most instances, especially cleanrooms. Not recommended for use as a design criterion, only for evaluation.1.56N/A
VC-GAppropriate for extremely quiet research spaces; generally difficult to achieve in most instances, especially cleanrooms. Not recommended for use as a design criterion, only for evaluation.0.78N/A

1. As measured in one-third octave bands of frequency over the frequency 8 to 80 Hz (VC-A and VC-B) or 1 to 80 Hz (VC-C through VC-G).

2. The detail size refers to line width in the case of microelectronics fabrication, the particle (cell) size in the case of medical and pharmaceutical research, etc.

3. The detail size is not relevant to imaging associated with probe technologies, AFMs, and nanotechnology.