HEM6000 SEM DVIA-MB1000 (S/N 250514R3) Installation Report
Contents
Overview
Following the installation of DVIA-MB1000 at Hangzhou, tuning and vibration measurement were performed.
Vibration measurement was conducted with the equipment in the IDLE state.
Data is represented in VC Curve format and reference material on the vibration level is provided.
Vibration Isolation System Information
Model: DVIA-MB1000
Serial Number: 250514R3
Engineer
Gyumin.park · DAEIL SYSTEMS
Tuning Date
2026.10.27
End User
Westlake University
Number of Tuning Trial
1st
Site
Hangzhou
Equipment
Manufacturer: CIQTEK
Equipment: SEM
Model: HEM6000
Vibration Specification
Horizontal
| Frequency [Hz] | Allowable [µm/s RMS] |
|---|---|
| 0 ~ 15 | 2 |
| 15 ~ 60 | 6 |
| Above 60 | 10 |
Vertical
| Frequency [Hz] | Allowable [µm/s RMS] |
|---|---|
| 0 ~ 15 | 2 |
| 15 ~ 60 | 8 |
| Above 60 | 10 |
Equipment Condition
Turn Off
Tuning Request
Measurement Summary
| Measurement Point | Vibration Specifications | Axis | Result | |
|---|---|---|---|---|
| Floor | DVIA-MB1000 | |||
| 1. Floor 2. DVIA-MB1000 | See Section 9 | Vertical | ✗ FAIL | ✓ PASS |
| Left to Right | ✓ PASS | ✓ PASS | ||
| Front to Back | ✗ FAIL | ✓ PASS | ||
Measurement Data Obtained via UI Program
Vertical — Autospectrum
Left to Right — Autospectrum
Front to Back — Autospectrum
Installation Photo
Equipment
Reference
The table below classifies vibration levels based on IEST-RP-CC012.2.
¹ As measured in one-third octave bands of frequency over the frequency 8 to 80 Hz (VC-A and VC-B) or 1 to 80 Hz (VC-C through VC-G).
² The detail size refers to the width in the case of microelectronics fabrication, the particle (cell) size in the case of medical and pharmaceutical research, etc. It is not relevant to imaging associated with probe technologies, AFMs, and nanotechnology.
The information given in this table is for guidance only. In most instances, it is recommended that the advice of someone knowledgeable about applications and vibration requirements of the equipment and processes be sought.