Installation Report
DVIA-M Series
03-20-2026
Laser Science DVIA-MO1000 (211126R3-5) Installation Report — March 20, 2026
DVIA-M Series
Installation Report
Laser Science
DVIA-MO1000
211126R3-5
Contents
Customer
Laser Science
Author
Engineer: 이채원
Installation date: 26.03.18
Report written date: 26.03.20
Overview
Tuning was performed following relocation installation of the DVIA-MO1000.
Tuning was performed with the equipment not installed.
Vibration Isolation System Information
Model: DVIA-MO1000
Serial Number: 211126R3-5
Engineer
DAEIL SYSTEMS field engineer 이채원
Tuning Date
March 18, 2026
Installation Site
N/A
End User
Laser Science
Customer Equipment
Manufacturer: -
Equipment: -
Model: -
Equipment Specification
N/A
Equipment Status
장비미설치
Conclusion
Floor vibration is attenuated by 90% or more in the 2–10 Hz band in all directions.
Data and Image
Vertical Transmissibility
Left to Right Transmissibility
Front to Back Transmissibility