Skip to main content
Installation Report
DVIA-M Series
03-20-2026

Laser Science DVIA-MO1000 (211126R3-2) Installation Report — March 20, 2026

DVIA-M Series
Installation Report
Laser Science
DVIA-MO1000
211126R3-2

Customer

Laser Science

Author

N/A

Engineer: Chaewon Lee

Installation date: 26.03.18

Report written date: 26.03.20

Overview

Tuning was performed following relocation installation of the DVIA-MO1000.

Tuning was carried out while the laser was not installed.

There was no output from the Z Base Sensor. Z-axis performance was not saved separately; performance from another product tuned previously was added to this report for delivery to the customer.

Vibration isolation platform

Model: DVIA-MO1000

Serial Number: 211126R3-2

Engineer

Chaewon Lee, Field Engineer, DAEIL SYSTEMS

Tuning Date

March 18, 2026

Installation site

N/A

End User

Laser Science

Customer equipment

Manufacturer: N/A

Equipment: N/A

Model: N/A

Equipment specification

N/A

Equipment status

Equipment not installed

Conclusion

In all directions, in the 2–10 Hz band, floor vibration is attenuated by 90% or more.

Data and images

Vertical Transmissibility

Left to Right Transmissibility

Front to Back Transmissibility

Z Base Sensor Output

Share this Case Study

Case Study Information

Category
Installation Report
SeriesDVIA-M Series
Date03-20-2026
Tags
DVIA-M Series
Installation Report
Laser Science
DVIA-MO1000
211126R3-2