Installation Report
DVIA-ULF Series
12-23-2020

POSTECH Scienta Omicron PEEM DVIA-U700 (P18) Replacement Installation Report — Pohang

DVIA-U Series
Installation Report
POSTECH
Pohang
Scienta Omicron
PEEM
DVIA-U700
P18
RIST

POSTECH U700 P18

포항공대_U700_ P18

Site checklist

Target equipmentScienta Omicron-IEF/TOF PEEM
Place of installationRist 본부 1층
RemarkGood Performance
END USER포항공대 물리학과

Operator and dates

Operator: Choi, Hyoung Mun Date of setup : 20.12.18 Date of reporting : 20.12.23Date of business trip : 17.12.27~17.12.29

Overview

Overview After Active mounting, measure the vibration environment to check that there is no problem in using the equipment. Measurement date: 2020.12.18 (Fri), Operator: Daeil System, Manager Choi, Hyoung Mun Place of measurement : Rist 본부 1층 Equipment Model: Scienta Omicron-IEF/TOF PEEM Measurement conditions and results - DAQ System : B&K PULSE 14 & LAN-XI Type 3056 - Accelerometer : PCB PIEZOTRONICS 393B05 - Frequency Range : 0.7 ~ 200Hz

Summary of measurement results

Test ConditionVC-Class
ZXY
FloorAEE
Above ActiveFFF

Good omnidirectional performance.

- Frequency Range : 0.7 ~ 200Hz

- Frequency Range : 0.7 ~ 200Hz

- Frequency Range : 0.7 ~ 200Hz

Good omnidirectional performance.

Good omnidirectional performance.

U700 internal program performance measurement result

Z axis

X axis

Y axis

Octave Band Result (BASE: Floor, TOP: Above Active)

Z axis

X axis

Y axis