Skip to main content
Installation Report
DVIA-M Series
09-17-2018

Zhejiang University HITACHI H-7650 TEM DVIA-MB1000 (MB1000 P70) Installation Report — 2018-09-17

DVIA-M Series
Installation Report
Zhejiang University
Hangzhou
HITACHI
TEM
H-7650
DVIA-MB1000
MB1000
P70
2018

(Zhejiang University MB1000_P70 Set up)

Target equipment

HITACH: H-7650 Transmission Electron Microscope (TEM)

Place of installation

Zhejiang University, 1th FLOOR

Uniqueness

Good Performance

END USER

Zhejiang University of Traditional Chinese Medicine

Date of business trip

Date of business trip : 18.09.12

Business traveler

Business traveler: Choi, HyoungMun

Date Created

Date Created : 18.09.17

Overview

After Active mounting, measure the vibration environment to check that there is no problem in using the equipment.

Measurement date

Measurement date: 2018.09.12 (Wednesday), Rater: Daeil System, Manager Choi , Hyoung Moon

Place of measurement

Place of measurement : Zhejiang University, 1th FLOOR

Equipment Model

Equipment Model: HITACH: H-7650 Transmission Electron Microscope (TEM)

Measurement conditions and results

N/A

Summary of measurement results

Test ConditionVC-Class ZVC-Class XVC-Class Y
FloorDEE
Above ActiveGGG

Good omnidirectional performance and attached VC-Class graph.

MB1000 internal program performance measurement result

Z axis - internal program

X axis - internal program

Y axis - internal program

Octave Band Result (BASE: Floor, TOP: Above Active)

Z axis - octave band result

X axis - octave band result

Y axis - octave band result