설치 보고서
DVIA-M Series
11-10-2023
POSCO FUTURE M GeminiSEM 560 DVIA-MB1000 (230704R3) Installation Report
DVIA-M Series
Installation Report
POSCO
ZEISS
GeminiSEM
DVIA-MB1000
230704R3
SEM
세종시
음극재
목차
Prepared for
POSCO FUTURE M Co., Ltd.
Engineer
Sangjae Lee, Jongwon Park
Tuning date
23.11.09
Report written date
23.11.10
Date of business trip
17.12.27~17.12.29
개요
The DVIA-MB1000 active vibration isolation platform is appropriately installed and functioning as intended.
시스템 정보
모델: DVIA-MB1000
시리얼 번호: 230704R3
Engineer
Sangjae Lee, Jongwon Park from DAEIL SYSTEMS
Tuning Date
November 09, 2023
설치 위치
POSCO FUTURE M 음극재 2공장 1층 SEM룸, 세종시
장비
GeminiSEM 560
진동 결과 요약
| Measurement Point | Z-axis (Vertical) 1–10 Hz | Z-axis (Vertical) 12.5–80 Hz | X-axis (Left to Right) 1–10 Hz | X-axis (Left to Right) 12.5–80 Hz | Y-axis (Front to Back) 1–10 Hz | Y-axis (Front to Back) 12.5–80 Hz |
|---|---|---|---|---|---|---|
| Floor | E | G | E | G | F | G |
| Active | G | G | G | G | G | G |
데이터 및 이미지
Z-axis (Vertical) Floor VC curves
Z-axis (Vertical) Active VC curves
X-axis (Left to Right) Floor VC curves
X-axis (Left to Right) Active VC curves
Y-axis (Front to Back) Floor VC curves
Y-axis (Front to Back) Active VC curves
장비 사진
Reference
일반 진동 기준
참고:
- VC-A/B는 8-80 Hz, VC-C~VC-G는 1-80 Hz 범위의 1/3 옥타브 밴드에서 측정.
- 세부 크기는 마이크로일렉트로닉스 제조에서의 선폭 또는 의료 연구에서의 입자 크기를 의미.
이 케이스 스터디 공유하기
케이스 스터디 정보
카테고리
설치 보고서
시리즈DVIA-M Series
작성일11-10-2023
태그
DVIA-M Series
Installation Report
POSCO
ZEISS
GeminiSEM
DVIA-MB1000
230704R3
SEM
세종시
음극재
관련 케이스 스터디

DVIA-M Series
YIMO ZEISS Gemini Sigma 300 DVIA-MB1000 Installation Report (MB1000 P155, 2020-05-08)
05-08-2020자세히

DVIA-M Series
Northwest Polytech University ZEISS GeminiSEM 300 DVIA-MB1000 Installation Report (190123R1, 2019-12-16)
12-16-2019자세히

DVIA-M Series
Northwest Polytechnical University ZEISS GeminiSEM300 DVIA-MB1000 (P115) trip report — 2019.12.16
12-16-2019자세히