본문으로 바로가기
설치 보고서
DVIA-M Series
06-29-2021

Nexchip Zeiss Crossbeam 550 DVIA-MB1000 (210204R3) Internal Program Performance — Hefei — 2021.06.29

DVIA-M Series
Installation Report
Nexchip
Hefei
ZEISS Crossbeam 550
DVIA-MB1000
210204R3

MB1000 P212 210204R3

Target equipmentZEISS-CROSSBEAM 550
Place of installationhefei
RemarkGood Performance
END USERNexchip
Operator: Choi, HyoungMunDate of business trip : 17.12.27~17.12.29
Date of setup : 21.06.28
Date of reporting : 21.06.29

개요

After Active mounting, measure the vibration environment to check that there is no problem in using the equipment.

Measurement date: 21.06.28 (MON)

Operator: Daeil System, Research engineer Choi, HyoungMun

Place of measurement : hefei

Equipment Model: ZEISS-CROSSBEAM 550

측정 조건 및 결과

측정 결과 요약

Transmissibility (2 Hz)Transmissibility (5 Hz)Transmissibility (10 Hz)
Z-23 dB-37 dB-37 dB
X-33 dB-30 dB-30 dB
Y-27 dB-34 dB-40 dB

Good omnidirectional performance.

MB1000 internal program performance measurement result

Z축

X축

Y축

이 케이스 스터디 공유하기

케이스 스터디 정보

카테고리
설치 보고서
시리즈DVIA-M Series
작성일06-29-2021
태그
DVIA-M Series
Installation Report
Nexchip
Hefei
ZEISS Crossbeam 550
DVIA-MB1000
210204R3