설치 보고서
DVIA-M Series
06-10-2021
YIMO remote tuning — Inner Mongolia Univ. Hitachi SU8100 DVIA-MB1000 (201216R1-2 · P211) — 2021.06.10
DVIA-M Series
Installation Report
YIMO
DVIA-MB1000
201216R1-2
P211
Hitachi
SU8100
Inner Mongolia University
remote
목차
(MB1000_201216R1-2(P211) Set up)
대상 장비
Hitachi SU8100
설치 장소
NeiMengGu
비고
Good Performance
END USER
内蒙古大学
Operator
Munho Noh
Date of setup
21.06.10
Date of reporting
21.06.10
Date of business trip
17.12.27~17.12.29
개요
After Active mounting, measure the vibration environment to check that there is no problem in using the equipment.
측정 일자
Measurement date: 21.06.10 (THU)
Operator: Daeil System, Research engineer Munho Noh
측정 장소
NeiMengGu
Equipment Model
Hitachi SU8100
측정 조건 및 결과
측정 결과 요약
Test Condition
Transmissibility
| 2 Hz | 5 Hz | 10 Hz | |
|---|---|---|---|
| Z | -24.20 dB | -33.99 dB | -30.13 dB |
| X | -27.78 dB | -35.01 dB | -39.27 dB |
| Y | -38.12 dB | -33.06 dB | -43.90 dB |
Good omnidirectional performance.
MB1000 internal program performance measurement result
Z axis
X axis
Y axis
이 케이스 스터디 공유하기
케이스 스터디 정보
카테고리
설치 보고서
시리즈DVIA-M Series
작성일06-10-2021
태그
DVIA-M Series
Installation Report
YIMO
DVIA-MB1000
201216R1-2
P211
Hitachi
SU8100
Inner Mongolia University
remote
관련 케이스 스터디

DVIA-M Series
YIMO Hitachi SU8100 MB1000 P261 (211013R1) internal program handout — 2021.12.13
12-13-2021자세히

DVIA-M Series
Inner Mongolia University Hitachi S4800 DVIA-MB1000 remote VEC (201216R1-1 / P209) — 2021.06.10
06-10-2021자세히

DVIA-M Series
YIMO Allide (Shanghai) Hitachi SU8600 FE-SEM DVIA-MB1000 (240103R6) Installation Report
05-11-2026자세히