본문으로 바로가기
설치 보고서
DVIA-M Series
09-10-2020

Shandong Lanxiang Zeiss Evo 18 DVIA-MB1000 (200522R7 · P163) Remote measurement — Internal program plots — 2020.09.10

DVIA-M Series
Installation Report
DVIA-MB1000
200522R7
Zeiss
Evo 18
Shandong
YIMO

(MB1000_200522R7(P163) Set up)

Target equipmentZeiss Evo 18
Place of installationSHANDONG
RemarkGood Performance
END USER山东蓝翔
Operator: Munho NohDate of business trip : 17.12.27~17.12.29
Date of setup : 20.09.10Date of reporting : 20.09.10

개요

After Active mounting, measure the vibration environment to check that there is no problem in using the equipment.

Measurement date: 20.09.10 (THU)

Operator: Daeil System, Research engineer Munho Noh

Place of measurement : Shangdong

Equipment Model: Zeiss EVO 18

측정 조건 및 결과

측정 결과 요약

Test Condition

Transmissibility

2 Hz5 Hz10 Hz
Z-24.07 dB-32.92 dB-27.44 dB
X-23.46 dB-21.65 dB-24.03 dB
Y-21.48 dB-23.73 dB-43.64 dB

Good omnidirectional performance.

MB1000 internal program performance measurement result

Z축

X축

Y축

이 케이스 스터디 공유하기

케이스 스터디 정보

카테고리
설치 보고서
시리즈DVIA-M Series
작성일09-10-2020
태그
DVIA-M Series
Installation Report
DVIA-MB1000
200522R7
Zeiss
Evo 18
Shandong
YIMO