본문으로 바로가기
설치 보고서
DVIA-M Series
08-17-2020

YIMO Zeiss G300 DVIA-MB1000 (200319R2 · P164) Remote measurement — Internal program plots — 2020.08.17

DVIA-M Series
Installation Report
DVIA-MB1000
200319R2
P164
Zeiss
CSOT
YIMO
Shenzen
remote tuning

(MB1000_200319R2(P164) Set up)

Target equipmentZeiss G300
Place of installationShenzen
RemarkGood Performance
END USERTCL华星光电技术有限公司
Operator: Munho NohDate of business trip : 17.12.27~17.12.29
Date of setup : 20.08.17Date of reporting : 20.08.17

개요

After Active mounting, measure the vibration environment to check that there is no problem in using the equipment.

Measurement date: 20.08.17 (MON)

Operator: Daeil System, Research engineer Munho Noh

Place of measurement : Shenzen

Equipment Model: Zeiss G300

측정 조건 및 결과

측정 결과 요약

2 Hz5 Hz10 Hz
Z-22.05 dB-25.11 dB-28.19 dB
X-25.72 dB-40.11 dB-27.52 dB
Y-28.21 dB-37.86 dB-41.08 dB

Good omnidirectional performance.

MB1000 internal program performance measurement result

Z축

X축

Y축

이 케이스 스터디 공유하기

케이스 스터디 정보

카테고리
설치 보고서
시리즈DVIA-M Series
작성일08-17-2020
태그
DVIA-M Series
Installation Report
DVIA-MB1000
200319R2
P164
Zeiss
CSOT
YIMO
Shenzen
remote tuning