설치 보고서
DVIA-M Series
02-03-2020
Seoul National University School of Dentistry Thermo Fisher Apreo S DVIA-MB1000 (MB1000_P157 / 190816R4) Installation Report
DVIA-M Series
Installation Report
Seoul National University School of Dentistry
Thermo Fisher
Apreo S
DVIA-MB1000
MB1000_P157
190816R4
목차
설치 현장 사진
개요
| Target equipment | THERMOFISHER-APREO S |
|---|---|
| Place of installation | F1, 서울대학교 치과대학원 |
| Remark | Good Performance |
| END USER | 서울대학교 치과대학원 |
Operator: Choi, Hyoung Mun
Date of setup : 20.01.31
Date of reporting : 20.02.03
Date of business trip : 17.12.27~17.12.29
After Active mounting, measure the vibration environment to check that there is no problem in using the equipment.
Measurement date: 2020.01.31 (FRI),
Operator: Daeil System, Manager Choi, Hyoung Mun
Place of measurement : F1, 서울대학교 치과대학원
Equipment Model: THERMOFISHER-APREO S
Model: DVIA-MB1000
Serial Number: MB1000_P157
측정 조건 및 결과
측정 결과 요약
| 시험 조건 | Z | X | Y |
|---|---|---|---|
| Floor | B | E | E |
| Above Active | G | E | G |
Good omnidirectional performance.
MB1000 내부 프로그램 성능 측정 결과
내부 프로그램 Z축
내부 프로그램 X축
내부 프로그램 Y축
Octave Band Result (BASE: Floor, TOP: Above Active)
옥타브 밴드 Z축
옥타브 밴드 X축
옥타브 밴드 Y축
이 케이스 스터디 공유하기
케이스 스터디 정보
카테고리
설치 보고서
시리즈DVIA-M Series
작성일02-03-2020
태그
DVIA-M Series
Installation Report
Seoul National University School of Dentistry
Thermo Fisher
Apreo S
DVIA-MB1000
MB1000_P157
190816R4
관련 케이스 스터디

DVIA-M Series
Dongguan Jinuo Thermo Fisher Helios 5 DVIA-MB1000 (230712R3-2) Installation Report
07-05-2024자세히

DVIA-M Series
University of North Dakota TFS HT Tank 200 kV DVIA-MB1000 VEC (220524R4) Installation Report — 2023.08.25
08-25-2023자세히

DVIA-M Series
Seoul National University Thermo Fisher Glacios TEM DVIA-MB1000 (MB1000_P218) Installation Report
02-22-2021자세히