본문으로 바로가기
설치 보고서
DVIA-M Series
02-22-2021

Seoul National University Thermo Fisher Glacios TEM DVIA-MB1000 (MB1000_P218) Installation Report

DVIA-M Series
Installation Report
DVIA-MB1000
MB1000_P218
Seoul National University
Thermo Fisher
Glacios
TEM

설치 현장 사진

개요

Target equipmentThermo Fisher_BS Glacios (TEM)
Place of installation150동 1층 유전공학연구소
RemarkGood Performance
END USER서울대학교

Operator: Choi, HyoungMun

Date of setup : 21.02.22

Date of reporting : 21.02.22

Date of business trip : 17.12.27~17.12.29

After Active mounting, measure the vibration environment to check that there is no problem in using the equipment.

Measurement date: 21.02.22 (MON)

Operator: Daeil System, Research engineer Choi, HyoungMun

Place of measurement : 150동 1층 유전공학연구소

Equipment Model: Thermo Fisher_BS Glacios (TEM)

Model: DVIA-MB1000

Serial Number: MB1000_P218

측정 조건 및 결과

DAQ System : B&K PULSE 14 & LAN-XI Type 3056

Accelerometer : PCB PIEZOTRONICS 393B05

Frequency Range : 0.7 ~ 200Hz

B&K 측정 VC (Z축)

B&K 측정 VC (X축)

B&K 측정 VC (Y축)

측정 결과 요약

시험 조건ZXY
FloorCDE
Above ActiveGGG

Good omnidirectional performance.

측정 요약 도면 1

측정 요약 도면 2

MB1000 내부 프로그램 성능 측정 결과

내부 프로그램 Z축

내부 프로그램 X축

내부 프로그램 Y축

※ 진동 기준 곡선

기준 곡선 Z축

기준 곡선 X축

기준 곡선 Y축

이 케이스 스터디 공유하기

케이스 스터디 정보

카테고리
설치 보고서
시리즈DVIA-M Series
작성일02-22-2021
태그
DVIA-M Series
Installation Report
DVIA-MB1000
MB1000_P218
Seoul National University
Thermo Fisher
Glacios
TEM