설치 보고서
DVIA-M Series
08-25-2023
University of North Dakota TFS HT Tank 200 kV DVIA-MB1000 VEC (220524R4) Installation Report — 2023.08.25
DVIA-M Series
Installation Report
University of North Dakota
VEC
DVIA-MB1000
220524R4
Thermo Fisher
HT Tank
Contents
의뢰 기관
University of North Dakota
작성자
N/A
개요
The DVIA-MB1000 active vibration isolation platform is appropriately installed and functioning as intended.
시스템 정보
Model: DVIA-MB1000 - 910 x 1140 mm
Serial Number: 220524R4
엔지니어
Youngha Lee, Jongwon Park from DAEIL SYSTEMS
튜닝 날짜
August 25, 2023
보고서 작성일
23.09.06
출장 날짜
17.12.27~17.12.29
측정 장소
University of North Dakota Whitmer Hall
장비
TFS HT Tank 200 kV
진동 결과 요약
| 측정 지점 | Z축(수직) 1-10 Hz | Z축(수직) 12.5-80 Hz | X축(좌우) 1-10 Hz | X축(좌우) 12.5-80 Hz | Y축(전후) 1-10 Hz | Y축(전후) 12.5-80 Hz |
|---|---|---|---|---|---|---|
| Floor | F | D | G | F | G | G |
| Active | G | G | G | G | G | G |
데이터 및 이미지
8.1 UI Program Transmissibility
Z-axis (Vertical)
X-axis (Left to Right)
Y-axis (Front to Back)
VEC Measurement
Z-axis (Vertical) Floor VC Curve
Z-axis (Vertical) Active VC Curve
X-axis (Left to Right) Floor VC Curve
X-axis (Left to Right) Active VC Curve
Y-axis (Front to Back) Floor VC Curve
Y-axis (Front to Back) Active VC Curve
Transmissibility Graph
Reference
일반 진동 기준
참고:
1. VC-A/B는 8-80 Hz의 1/3 옥타브 밴드에서, VC-C부터 VC-G는 1-80 Hz에서 측정됩니다.
2. 상세 크기는 마이크로일렉트로닉스 제조의 선폭 또는 의료 연구의 입자 크기를 의미합니다.
이 케이스 스터디 공유하기
케이스 스터디 정보
카테고리
설치 보고서
시리즈DVIA-M Series
작성일08-25-2023
태그
DVIA-M Series
Installation Report
University of North Dakota
VEC
DVIA-MB1000
220524R4
Thermo Fisher
HT Tank
관련 케이스 스터디

DVIA-M Series
University of North Dakota Thermo Fisher Talos F200X DVIA-MB3000 VEC (221104R3) Installation Report — 2023.08.25
08-25-2023자세히

DVIA-M Series
YIMO Changping P3 Laboratory Thermo Fisher HT Tank DVIA-MB1000 (250708R5-2) Installation Report
11-11-2025자세히

DVIA-M Series
Dongguan Jinuo Thermo Fisher Helios 5 DVIA-MB1000 (230712R3-2) Installation Report
07-05-2024자세히