KRISS FE SEM UB700 Vibration Measurement Report (2016-06-21)
DVIA-U Series
Installation Report
Vibration Measurement
KRISS
Korea Research Institute of Standards and Science
FE SEM
UB700
VC-Class
2016
Jongwook Lim
Contents
Target equipment
FE SEM in-house developed model
Author and report date
Author: Jongwook Lim, Assistant Manager, DAEIL SYSTEMS
Report written date: 2016.06.21
Installation result summary
| Test Direction | VC-Class | ||
|---|---|---|---|
| Base | Top | ||
| ACTIVE | Z | C | E |
| ACTIVE | X | E | G |
| ACTIVE | Y | E | G |
For the vertical axis, performance is good from C Class to E Class; for the horizontal axes, performance is good from E Class to G Class respectively.
Measurement sensors were placed as shown in the figure below to measure vibration of the laboratory floor (BASE) and the top of the vibration isolation platform (TOP).
Measurement setup diagram
Vibration environment measurement result
Z axis
X axis
Y axis