Thermo Fisher Scientific Apreo 2S HiVac DVIA-ML1000 Installation Report
Contents
Overview
Following the installation of DVIA-ML1000 at Wuhan, tuning and vibration measurement were performed.
Vibration measurement was conducted with the equipment in the Idle state.
Data is represented in VC Curve format and reference material on the vibration level is provided.
Vibration Isolation System Information
Model: DVIA-ML1000
Serial Number: 260730R4
Engineer
Taehun KIM · DAEIL SYSTEMS
Tuning Date
2026.10.20
End User
Wuhan University of Science and Technology
Number of Tuning Trial
1st
Site
Wuhan
Equipment
Manufacturer: Thermo Fisher Scientific
Equipment: SEM
Model: Apreo 2S HiVac
Vibration Specification
VC-F
Equipment Condition
Idle
Tuning Request
Conclusion
Following installation and tuning of the DVIA-ML1000 (S/N 260730R4) at Wuhan University of Science and Technology, vibration was measured on the floor and on the isolator with the Thermo Fisher Scientific Apreo 2S HiVac in the Idle state.
The required vibration specification for the equipment is VC-F. As shown in Section 13, the floor vibration exceeded VC-F in all three axes. The DVIA-ML1000 reduced it to within VC-F in all three axes (Vertical, Left to Right, Front to Back).
The DVIA-ML1000 therefore provides a vibration environment that meets the requirements of the Apreo 2S HiVac, and no additional tuning or corrective action is required at this time.
Measurement Summary
| Measurement Point | Vibration Specifications | Axis | Result | |
|---|---|---|---|---|
| Floor | DVIA-ML1000 | |||
| 1. Floor 2. DVIA-ML1000 | See Section 9 | Vertical | ✗ FAIL | ✓ PASS |
| Left to Right | ✗ FAIL | ✓ PASS | ||
| Front to Back | ✗ FAIL | ✓ PASS | ||
Measurement Data Obtained via UI Program
Vertical — VC Curves
Left to Right — VC Curves
Front to Back — VC Curves
Installation Photo
Equipment
Reference
The table below classifies vibration levels based on IEST-RP-CC012.2.
Notes:
1. As measured in one-third octave bands of frequency over the frequency 8 to 80 Hz (VC-A and VC-B) or 1 to 80 Hz (VC-C through VC-G).
2. The detail size refers to the width in the case of microelectronics fabrication, the particle (cell) size in the case of medical and pharmaceutical research, etc. It is not relevant to imaging associated with probe technologies, AFMs, and nanotechnology.
The information given in this table is for guidance only. In most instances, it is recommended that the advice of someone knowledgeable about applications and vibration requirements of the equipment and processes be sought.