설치 보고서
DVIA-M Series
09-13-2021
YIMO STM OEM DVIA-MO1000 (P247) Second Setup Installation Report — September 13, 2021
DVIA-M Series
Installation Report
YIMO
STM
DVIA-MO1000
P247
MO1000 internal program
목차
(MO1000 P247 2ND SETUP)
| 설치 장비(Target equipment) | STM OEM |
|---|---|
| 설치 위치 | N/A |
| 비고(Remark) | Good Performance |
| END USER | N/A |
Operator: Choi, HyoungMun
Date of setup : 21.09.10
Date of reporting : 21.09.13 | Date of business trip : 17.12.27~17.12.29
개요
After Active mounting, measure the vibration environment to check that there is no problem in using the equipment.
Measurement date: 21.09.10 (FRI)
Operator: Daeil System, Research engineer Choi, HyoungMun
Place of measurement : N/A
Equipment Model: STM OEM
측정 조건 및 결과
측정 결과 요약
Good omnidirectional performance.
| 2 Hz | 5 Hz | 10 Hz | |
|---|---|---|---|
| Z | -28 dB | -27 dB | -29 dB |
| X | -30 dB | -20 dB | -28 dB |
| Y | -30 dB | -30 dB | -28 dB |
MO1000 internal program performance measurement result
Z axis
X axis
Y axis
Reference
Generic Vibration Criteria
참고:
1. VC-A/B는 8-80 Hz, VC-C~VC-G는 1-80 Hz 범위의 1/3 옥타브 밴드에서 측정.
2. 세부 크기는 마이크로일렉트로닉스 제조에서의 선폭 또는 의료 연구에서의 입자 크기를 의미합니다.