설치 보고서
DVIA-ULF Series
12-23-2020
POSTECH Scienta Omicron PEEM DVIA-U700 (P18) Replacement Installation Report — Pohang
DVIA-U Series
Installation Report
POSTECH
Pohang
Scienta Omicron
PEEM
DVIA-U700
P18
RIST
목차
POSTECH U700 P18
포항공대_U700_ P18
현장 확인 항목
| Target equipment | Scienta Omicron-IEF/TOF PEEM |
|---|---|
| Place of installation | Rist 본부 1층 |
| Remark | Good Performance |
| END USER | 포항공대 물리학과 |
운영자 및 일정
| Operator: Choi, Hyoung Mun Date of setup : 20.12.18 Date of reporting : 20.12.23 | Date of business trip : 17.12.27~17.12.29 |
|---|
개요
Overview After Active mounting, measure the vibration environment to check that there is no problem in using the equipment. Measurement date: 2020.12.18 (Fri), Operator: Daeil System, Manager Choi, Hyoung Mun Place of measurement : Rist 본부 1층 Equipment Model: Scienta Omicron-IEF/TOF PEEM Measurement conditions and results - DAQ System : B&K PULSE 14 & LAN-XI Type 3056 - Accelerometer : PCB PIEZOTRONICS 393B05 - Frequency Range : 0.7 ~ 200Hz
측정 결과 요약
| Test Condition | VC-Class | ||
|---|---|---|---|
| Z | X | Y | |
| Floor | A | E | E |
| Above Active | F | F | F |
Good omnidirectional performance.
- Frequency Range : 0.7 ~ 200Hz
- Frequency Range : 0.7 ~ 200Hz
- Frequency Range : 0.7 ~ 200Hz
Good omnidirectional performance.
Good omnidirectional performance.
U700 내부 프로그램 성능 측정 결과
Z axis
X axis
Y axis
Octave Band Result (BASE: Floor, TOP: Above Active)
Z axis
X axis
Y axis
이 케이스 스터디 공유하기
케이스 스터디 정보
카테고리
설치 보고서
시리즈DVIA-ULF Series
작성일12-23-2020
태그
DVIA-U Series
Installation Report
POSTECH
Pohang
Scienta Omicron
PEEM
DVIA-U700
P18
RIST
관련 케이스 스터디

DVIA-ULF Series
Liupan Mountain Laboratory ZEISS Sigma 360 FE-SEM DVIA-U700 (241115R3) Installation Report — Yinchuan
07-14-2025자세히

DVIA-ULF Series
YIMO DVIA-UD350 (210929R1, P26) OLIMPUS remote tuning worksheet — UD350 internal Z / X / Y program plots
05-26-2026자세히

DVIA-ULF Series
KRISS FE SEM UB700 Vibration Measurement Report (2016-06-21)
05-26-2026자세히