설치 보고서
DVIA-P Series
05-26-2026

Xiamen University of Technology Zeiss Sigma 500 UB700 Installation Report (181019R7)

DVIA-P Series
Installation Report
Xiamen University of Technology
Zeiss
Sigma 500
UB700
181019R7
VC-Class
Floor vs Active
Octave band

대상 장비

Zeiss Sigma 500

설치 장소

SEM Lab, 1F, Xiamen University of Technology

비고

Good performance

END USER

Xiamen University of Technology

Operator: Choi, Sung Won

Date of setup : 19.01.03

Date of reporting : 19.01.07

Date of business trip : 17.12.27~17.12.29

개요

After Active mounting, measure the vibration environment to check that there is no problem in using the equipment.

Measurement date: 2019.01.03 (Tuesday),

Operator: Daeil System, Manager Sung Won Choi

Place of measurement : SEM Lab, 1st Floor, Xiamen University of Technology

Equipment Model: Zeiss Sigma 500

측정 조건 및 결과

측정 결과 요약

Test Condition: VC-ClassZXY
FloorCDD
Above ActiveCFE

Good omnidirectional performance.

In all direction, vibration level is high at 100Hz. Acoustic or direct force can be the cause.

MB1000 내장 프로그램 성능 측정 결과

Z축

X축

Y축

Octave Band Result (BASE: Floor, TOP: Above Active)

Z축

X축

Y축