Xiamen University of Technology Zeiss Sigma 500 UB700 Installation Report (181019R7)
DVIA-P Series
Installation Report
Xiamen University of Technology
Zeiss
Sigma 500
UB700
181019R7
VC-Class
Floor vs Active
Octave band
목차
대상 장비
Zeiss Sigma 500
설치 장소
SEM Lab, 1F, Xiamen University of Technology
비고
Good performance
END USER
Xiamen University of Technology
Operator: Choi, Sung Won
Date of setup : 19.01.03
Date of reporting : 19.01.07
Date of business trip : 17.12.27~17.12.29
개요
After Active mounting, measure the vibration environment to check that there is no problem in using the equipment.
Measurement date: 2019.01.03 (Tuesday),
Operator: Daeil System, Manager Sung Won Choi
Place of measurement : SEM Lab, 1st Floor, Xiamen University of Technology
Equipment Model: Zeiss Sigma 500
측정 조건 및 결과
측정 결과 요약
| Test Condition: VC-Class | Z | X | Y |
|---|---|---|---|
| Floor | C | D | D |
| Above Active | C | F | E |
Good omnidirectional performance.
In all direction, vibration level is high at 100Hz. Acoustic or direct force can be the cause.
MB1000 내장 프로그램 성능 측정 결과
Z축
X축
Y축
Octave Band Result (BASE: Floor, TOP: Above Active)
Z축
X축
Y축