본문으로 바로가기
설치 보고서
DVIA-M Series
08-19-2020

Zhejiang Zili Co., Ltd. Zeiss Sigma 500 DVIA-MB1000 (200715R3-1 · P186) Remote measurement — Internal program plots — 2020.08.19

DVIA-M Series
Installation Report
DVIA-MB1000
200715R3-1
Zeiss
Sigma 500
Thermo Scientific
Shaoxing
YIMO
P186

(MB1000_200715R3-1(P186) Set up)

Target equipmentZeiss sigma 500
Place of installationShaoxing,3rd floor
RemarkGood Performance
END USER浙江自立股份有限公司,
Operator: Noh, MunhoDate of business trip : 17.12.27~17.12.29
Date of setup : 20.08.19Date of reporting : 20.08.19

개요

After Active mounting, measure the vibration environment to check that there is no problem in using the equipment.

Measurement date: 20.08.19 (WED),

Operator: Daeil System, Manager Munho, Noh

Place of measurement : Shaoxing,自立股份有限公司,3rd floor

Equipment Model: Thermo scientific Scios 2

측정 조건 및 결과

측정 결과 요약

2 Hz5 Hz10 Hz
Z-24.25 dB-27.05 dB-31.71 dB
X-27.22 dB-35.68 dB-38.90 dB
Y-24.92 dB-37.14 dB-38.4d dB

Good omnidirectional performance.

MB1000 internal program performance measurement result

Z축

X축

Y축

이 케이스 스터디 공유하기