설치 보고서
DVIA-M Series
05-12-2020
ZEISS GeminiSEM300 DVIA-MB1000 Program (P115 / MB1000_190123R1) Installation Report — 2020.05.12
DVIA-M Series
Installation Report
ZEISS
GeminiSEM300
DVIA-MB1000
P115
MB1000 Program
190123R1
Northwest Polytech University
목차
(MB1000_190123R1 Set up)
| Target equipment | ZEISS GeminiSEM300 |
|---|---|
| Place of installation | 1F 126 SEM&TEM Room, Qixiang Building, Northwest Polytech University |
| Remark | Good Performance |
| END USER | Northwest Polytech University |
| Operator: Choi, HyoungMun | Date of business trip : 17.12.27~17.12.29 |
| Date of setup : 20.05.11 | |
| Date of reporting : 20.05.12 |
개요
After Active mounting, measure the vibration environment to check that there is no problem in using the equipment.
측정 일자
Measurement date: 20.05.11 (Mon),
Operator
Operator: Daeil System, Manager HyoungMun, Choi
측정 장소
Place of measurement : 1st Floor, Qixiang Building, Northwest Polytech University
Equipment Model
Equipment Model: ZEISS GeminiSEM300
측정 조건 및 결과
N/A
측정 결과 요약
| Test Condition | VC-Class Z | VC-Class X | VC-Class Y |
|---|---|---|---|
| Floor | N/A | N/A | N/A |
| Above Active | N/A | N/A | N/A |
Good omnidirectional performance.
MB1000 internal program performance measurement result
Z axis
X axis
Y axis
이 케이스 스터디 공유하기
케이스 스터디 정보
카테고리
설치 보고서
시리즈DVIA-M Series
작성일05-12-2020
태그
DVIA-M Series
Installation Report
ZEISS
GeminiSEM300
DVIA-MB1000
P115
MB1000 Program
190123R1
Northwest Polytech University
관련 케이스 스터디

DVIA-M Series
Northwest Polytech University ZEISS GeminiSEM 300 DVIA-MB1000 Installation Report (190123R1, 2019-12-16)
12-16-2019자세히

DVIA-M Series
Northwest Polytechnical University ZEISS GeminiSEM300 DVIA-MB1000 (P115) trip report — 2019.12.16
12-16-2019자세히

DVIA-M Series
ZEISS Sigma 300 VP DVIA-MB1000 (220121R3) Shanghai Installation Report
05-16-2026자세히