설치 보고서
DVIA-M Series
12-09-2019
BOSCH TESCAN-MIRA3 DVIA-MB1000 (MB1000 P141 · 190702R5) Internal program worksheet — 2019.12.09
DVIA-M Series
Installation Report
BOSCH
TESCAN
MIRA3
DVIA-MB1000
MB1000
P141
190702R5
목차
BOSCH
(MB1000_ 190702R5 Set up)
Target equipment
TESCAN-MIRA3
Place of installation
F2, BOSCH
Remark
Good Performance
END USER
BOSCH
Operator
Operator: Choi, Hyoung Mun
Date of setup
Date of setup : 19.12.05
Date of reporting
Date of reporting : 19.12.09
Date of business trip
Date of business trip : 17.12.27~17.12.29
개요
After Active mounting, measure the vibration environment to check that there is no problem in using the equipment.
측정 일자
Measurement date: 2019.12.05 (Thu),
Operator: Daeil System, Manager Choi, Hyoung Mun
측정 장소
Place of measurement : F2, BOSCH
Equipment Model
Equipment Model: TESCAN-MIRA3
측정 조건 및 결과
N/A
측정 결과 요약
| Test Condition | VC-Class Z | VC-Class X | VC-Class Y |
|---|---|---|---|
| Floor | D | E | E |
| Above Active | G | G | G |
Good omnidirectional performance.
MB1000 internal program performance measurement result
Z axis - internal program
X axis - internal program
Y axis - internal program
Octave Band Result (BASE: 바닥, TOP: Above Active)
Z axis - octave band result
X axis - octave band result
Y axis - octave band result
이 케이스 스터디 공유하기
케이스 스터디 정보
카테고리
설치 보고서
시리즈DVIA-M Series
작성일12-09-2019
태그
DVIA-M Series
Installation Report
BOSCH
TESCAN
MIRA3
DVIA-MB1000
MB1000
P141
190702R5
관련 케이스 스터디
DVIA-M Series
Tier-1 Semiconductor TESCAN SEM (MIRA 3) DVIA-MB1000 (MB1000 P39) vibration environment measurement — 2017.01.18
01-20-2017자세히
DVIA-M Series
LB Semicon TESCAN MIRA DVIA-MB1000 (MB1000 P220) Installation Report — 2021.04.16
04-16-2021자세히
DVIA-M Series
Tier-1 Semiconductor TESCAN-FE SEM DVIA-MB1000 (MB1000 program) Installation Report — 2020.06.29
06-29-2020자세히