저지앙 대학교 Hitachi SU8010 DVIA-MB1000(P68) MB1000 프로그램 워크시트 — 2018.09.17
DVIA-M Series
Installation Report
DVIA-MB1000
MB1000
P68
Hitachi
SEM
Zhejiang University
Hangzhou
Zhejiang University of Traditional Chinese Medicine
목차
Zhejiang University MB1000_P68 Set up
| Target equipment | HITACH-SUS8010 |
|---|---|
| Place of installation | Zhejiang University, 1th FLOOR |
| Uniqueness | Good Performance |
| END USER | Zhejiang University of Traditional Chinese Medicine |
| Business traveler: Choi, HyoungMun | Date of business trip : 18.09.12 |
| Date Created : 18.09.17 |
개요
After Active mounting, measure the vibration environment to check that there is no problem in using the equipment.
Measurement date: 2018.09.12 (Wednesday), Rater: Daeil System, Manager Choi , Hyoung Moon
Place of measurement : Zhejiang University, 1th FLOOR
Equipment Model: HITACH-SUS8010
측정 조건 및 결과
측정 결과 요약
| Test Condition | VC-Class Z | VC-Class X | VC-Class Y |
|---|---|---|---|
| Floor | C | E | E |
| Above Active | G | G | G |
Good omnidirectional performance and attached VC-Class graph.
MB1000 internal program performance measurement result
Z axis - internal program
X axis - internal program
Y axis - internal program
Octave Band Result (BASE: Floor, TOP: Above Active)
Z axis - octave band result
X axis - octave band result
Y axis - octave band result
Reference
Generic Vibration Criteria
참고:
1. VC-A/B는 8–80 Hz 1/3옥타브 밴드, VC-C–G는 1–80 Hz 1/3옥타브 밴드에서 측정한 값입니다.
2. Detail size는 마이크로전자 소자의 선폭이나 의료·바이오 분야에서는 입자 크기 등을 가리킵니다.