절강대학 HITACHI H-7650 TEM DVIA-MB1000(MB1000 P70) 설치 보고서 — 2018-09-17
DVIA-M Series
Installation Report
Zhejiang University
Hangzhou
HITACHI
TEM
H-7650
DVIA-MB1000
MB1000
P70
2018
목차
(Zhejiang University MB1000_P70 Set up)
Target equipment
HITACH: H-7650 Transmission Electron Microscope (TEM)
Place of installation
Zhejiang University, 1th FLOOR
Uniqueness
Good Performance
END USER
Zhejiang University of Traditional Chinese Medicine
Date of business trip
Date of business trip : 18.09.12
Business traveler
Business traveler: Choi, HyoungMun
Date Created
Date Created : 18.09.17
개요
After Active mounting, measure the vibration environment to check that there is no problem in using the equipment.
측정 일자
Measurement date: 2018.09.12 (Wednesday), Rater: Daeil System, Manager Choi , Hyoung Moon
측정 장소
Place of measurement : Zhejiang University, 1th FLOOR
Equipment Model
Equipment Model: HITACH: H-7650 Transmission Electron Microscope (TEM)
측정 조건 및 결과
N/A
측정 결과 요약
| Test Condition | VC-Class Z | VC-Class X | VC-Class Y |
|---|---|---|---|
| Floor | D | E | E |
| Above Active | G | G | G |
Good omnidirectional performance and attached VC-Class graph.
MB1000 internal program performance measurement result
Z axis - internal program
X axis - internal program
Y axis - internal program
Octave Band Result (BASE: Floor, TOP: Above Active)
Z axis - octave band result
X axis - octave band result
Y axis - octave band result