본문으로 바로가기
설치 보고서
DVIA-M Series
05-14-2018

East China University of Science and Technology ZEISS Gemini500 DVIA-MB1000 (MB1000 P63) Internal program worksheet — 2018.05.14

DVIA-M Series
Installation Report
East China University of Science and Technology
ZEISS
Gemini500
DVIA-MB1000
MB1000
P63

(MB1000_P63 Set up)

설치 사진

Target equipmentZEISS-Gemini500
Place of installationEast China University of Science and Technology
RemarksGood performance
END USEREast China University of Science and Technology
Operator: Choi, Sung WonDate of setup : 18.05.09
Date of reporting : 18.05.14

개요

After Active mounting, measure the vibration environment to check that there is no problem in using the equipment.

측정 일자

Measurement date: 2018.05.09 (Wednesday),

Operator

Operator: Daeil System, Manager Sung Won Choi

측정 장소

Place of measurement : 1th Floor, East China Univercity of Science and Technology

Equipment Model

Equipment Model: ZEISS-Gemini500

측정 조건 및 결과

측정 결과 요약

Test ConditionVC-Class ZVC-Class XVC-Class Y
FloorDEE
Above ActiveGGG

Good omnidirectional performance and attached VC-Class graph.

MB1000 internal program performance measurement result

Z axis - internal program

X axis - internal program

Y axis - internal program

Octave Band Result (BASE: Floor, TOP: Above Active)

Z axis - octave band result

X axis - octave band result

Y axis - octave band result

이 케이스 스터디 공유하기

케이스 스터디 정보

카테고리
설치 보고서
시리즈DVIA-M Series
작성일05-14-2018
태그
DVIA-M Series
Installation Report
East China University of Science and Technology
ZEISS
Gemini500
DVIA-MB1000
MB1000
P63