설치 보고서
DVIA-M Series
01-07-2022
ZEISS Sigma 300 DVIA-MB1000 (210503R7) YIMO internal program — 2022-01-07
DVIA-M Series
Installation Report
YIMO
ZEISS
Sigma 300
DVIA-MB1000
210503R7
MB1000
목차
(MB1000 P227 YIMO 210503R7)
| Target equipment | Zeiss Sigma300 |
|---|---|
| Place of installation | N/A |
| Remark | N/A |
| END USER | N/A |
| Operator: Choi, HyoungMun | Date of business trip : 17.12.27~17.12.29 |
| Date of setup : 22.01.04 | |
| Date of reporting : 22.01.07 |
개요
After Active mounting, measure the vibration environment to check that there is no problem in using the equipment.
Measurement date: 22.01.04 (tue)
Operator: Daeil System, Research engineer Choi, HyoungMun
Place of measurement : N/A
Equipment Model: Zeiss Sigma300
측정 조건 및 결과
측정 결과 요약
| Test condition | 2 Hz | 5 Hz | 10 Hz |
|---|---|---|---|
| Z | -26 dB | -35 dB | -34 dB |
| X | -38 dB | -37 dB | -35 dB |
| Y | -30 dB | -37 dB | -37 dB |
Good omnidirectional performance.
MB1000 internal program performance measurement result
Z축
X축
Y축
이 케이스 스터디 공유하기
케이스 스터디 정보
카테고리
설치 보고서
시리즈DVIA-M Series
작성일01-07-2022
태그
DVIA-M Series
Installation Report
YIMO
ZEISS
Sigma 300
DVIA-MB1000
210503R7
MB1000