본문으로 바로가기
설치 보고서
DVIA-M Series
12-29-2021

Baosteel ZEISS Sigma 300 DVIA-MB1000 (P205) Internal Program Performance — 2021.12.29

DVIA-M Series
Installation Report
Baosteel
ZEISS
Sigma 300
DVIA-MB1000
P205
MB1000

(MB1000 P205 Set up)

Target equipmentZEISS-Sigma300
Place of installationR&D Center 3Floor
RemarkGood Performance
END USERBAOSTELL
Operator: Choi, HyoungMunDate of business trip : 17.12.27~17.12.29
Date of setup : 21.12.29Date of reporting : 21.12.29

개요

After Active mounting, measure the vibration environment to check that there is no problem in using the equipment.

측정 일자

Measurement date: 21.12.29 (wed)

Operator

Operator: Daeil System, Research engineer Choi, HyoungMun

측정 장소

Place of measurement : R&D Center 3Floor

Equipment Model

Equipment Model: ZEISS-Sigma300

측정 조건 및 결과

측정 결과 요약

시험 조건2 Hz5 Hz10 Hz
Z-23 dB-40 dB-40 dB
X-27 dB-32 dB-35 dB
Y-30 dB-41 dB-40 dB

Good omnidirectional performance.

MB1000 internal program performance measurement result

Z축

X축

Y축