푸단대학교 MC1000(P76) Bruker UltraShield 500 NMR 성능 측정 보고서 — 2018-04-10
DVIA-M Series
Performance report
MC1000
P76
Bruker
NMR
Fudan University
DAEIL SYSTEMS
목차
Fudan University (MC1000_P76 Set up)
| Target equipment | BRUKER ULTRA SHIELD 500 NMR |
|---|---|
| Place of installation | Fudan UNIV, Basement 1FLOOR |
| Uniqueness | Good Performance |
| END USER | Fudan University |
| Date of business trip : 18.03.30 | Business traveler: Choi, HyoungMun |
| Date Created : 18.04.10 |
개요
After Active mounting, measure the vibration environment to check that there is no problem in using the equipment.
측정 정보
Measurement date: 2018.03.30 (Friday), Rater: Daeil System, Manager Choi , Hyoung Moon
Place of measurement : Fudan UNIV, Basement 1FLOOR
Equipment Model: BRUKER ULTRA SHIELD 500 NMR
측정 조건 및 결과
Summary of measurement results
| Test Condition | Z | X | Y |
|---|---|---|---|
| Floor | D | E | E |
| Above Active | G | G | G |
Good omnidirectional performance and attached VC-Class graph.
VC-Class graph
MB1000 internal program performance measurement result
Z axis - internal program
X axis - internal program
Y axis - internal program
Octave Band Result (BASE: Floor, TOP: Above Active)
Z axis - octave band result
X axis - octave band result
Y axis - octave band result
Reference
Generic Vibration Criteria
참고:
1. VC-A/B는 8–80 Hz 1/3옥타브 밴드, VC-C–G는 1–80 Hz 1/3옥타브 밴드에서 측정한 값입니다.
2. Detail size는 마이크로전자 소자의 선폭이나 의료·바이오 분야에서는 입자 크기 등을 가리킵니다.
이 케이스 스터디 공유하기
케이스 스터디 정보
카테고리
설치 보고서
작성일04-10-2018
태그
DVIA-M Series
Performance report
MC1000
P76
Bruker
NMR
Fudan University
DAEIL SYSTEMS
관련 케이스 스터디

DVIA-M Series
Fudan University BRUKER ULTRA SHIELD 500 NMR MC1000/P76 — DVIA-M 시리즈 성능 보고서(2018-04-10)
04-10-2018자세히

DVIA-M Series
YIMO 푸단대학교 현장 MC3000·MC1000·MB1000 제진 셋업 및 진동 측정 보고서 (2018-01-30)
01-30-2018자세히

DVIA-M Series
중국 상해 푸단(复旦)대학교 Bruker Ascend400(NMR) DVIA-MC1000(MC1000 P65 · 171114R5) — YIMO 셋업·측정 — 2018-01
01-26-2018자세히