설치 보고서
DVIA-M Series
04-10-2018
ZEISS SIGMA 300VP DVIA-MB1000 (MB1000 P28) 2nd Inspection Report — 2018.04.10
DVIA-M Series
Installation Report
DVIA-MB1000
MB1000
P28
ZEISS
SIGMA 300VP
DONGJI UNIV
목차
(DONGJI UNIV (MB1000_P28 Set up))
Target equipment
| Target equipment | ZEISS- SIGMA 300VP |
|---|---|
| Place of installation | DONGJI UNIV, 1FLOOR |
| Uniqueness | Good Performance |
| END USER | DONGJI UNIV |
| Date of business trip : 18.03.30 | |
| Business traveler: Choi, HyoungMun | |
| Date Created : 18.04.10 |
개요
After Active mounting, measure the vibration environment to check that there is no problem in using the equipment.
측정 일자
Measurement date: 2018.03.30 (Friday), Rater: Daeil System, Manager Choi , Hyoung Moon
측정 장소
Place of measurement : 1st Floor, DONGJI UNIV
Equipment Model
Equipment Model: ZEISS- SIGMA 300VP
측정 조건 및 결과
N/A
측정 결과 요약
| Test Condition | VC-Class Z | VC-Class X | VC-Class Y |
|---|---|---|---|
| Floor | E | E | E |
| Above Active | G | G | G |
Good omnidirectional performance and attached VC-Class graph.
MB1000 internal program performance measurement result
Z axis - internal program
X axis - internal program
Y axis - internal program
Octave Band Result (BASE: Floor, TOP: Above Active)
Z axis - octave band result
X axis - octave band result
Y axis - octave band result
Reference
Generic Vibration Criteria
참고:
1. VC-A/B는 8–80 Hz, VC-C~G는 1–80 Hz의 1/3 옥타브 밴드에서 측정된 값입니다.
2. Detail size는 반도체 제조의 폭 또는 의학 연구의 입자 크기를 가리킵니다.
이 케이스 스터디 공유하기
케이스 스터디 정보
카테고리
설치 보고서
시리즈DVIA-M Series
작성일04-10-2018
태그
DVIA-M Series
Installation Report
DVIA-MB1000
MB1000
P28
ZEISS
SIGMA 300VP
DONGJI UNIV
관련 케이스 스터디

DVIA-M Series
ZEISS Sigma 300 DVIA-MB1000 (210503R7) YIMO internal program — 2022-01-07
01-07-2022자세히

DVIA-M Series
ZEISS Sigma 300 DVIA-MB1000 (P227 / 210503R7) NEMO worksheet — 2022-01-07
01-07-2022자세히

DVIA-M Series
Baosteel ZEISS Sigma 300 DVIA-MB1000 (P205) Internal Program Performance — 2021.12.29
12-29-2021자세히