Hanyang University DVIA-MB1000 (141211R5) Installation Report — 2022-04-21
DVIA-M Series
Installation Report
Hanyang University
DVIA-M1000
DVIA-MB1000
141211R5
SEM
목차
Prepared for
Hanyang University
Prepared by
Engineer: Youngha Lee
Tuning date: 22.04.21
Report written date: 22.04.21
Date of business trip : 17.12.27~17.12.29
Overview
The DVIA-MB1000 active vibration isolation platform is appropriately installed and functioning as intended.
System Information
Model: DVIA-MB1000
Serial Number: 141211R5
Engineer
Youngha Lee, Seounghoon Cheon from DAEIL SYSTEMS
Tuning Date
April 21, 2022
Location
Hanyang University ,Sungdong Seoul
Equipment
SEM Equipment
Summary of Vibration Results
| Measurement Point | Z-axis (Vertical) 1-10 Hz | Z-axis (Vertical) 10-80 Hz | X-axis (Left to Right) 1-10 Hz | X-axis (Left to Right) 10-80 Hz | Y-axis (Front to Back) 1-10 Hz | Y-axis (Front to Back) 10-80 Hz |
|---|---|---|---|---|---|---|
| Floor | C | B | E | F | F | E |
| Active | G | E | G | G | G | G |
Data and Image
Z-axis (Vertical) VC curves
Floor
Active
Z-axis (Vertical) Transmissibility
X-axis (Left to Right) VC curves
Floor
Active
X-axis (Left to Right) Transmissibility
Y-axis (Front to Back) VC curves
Floor
Active
Y-axis (Front to Back) Transmissibility
Reference
일반 진동 기준
참고:
- VC-A/B는 8-80 Hz, VC-C~VC-G는 1-80 Hz 범위의 1/3 옥타브 밴드에서 측정.
- 세부 크기는 마이크로일렉트로닉스 제조에서의 선폭 또는 의료 연구에서의 입자 크기를 의미.
이 케이스 스터디 공유하기
케이스 스터디 정보
카테고리
설치 보고서
시리즈DVIA-M Series
작성일04-21-2022
태그
DVIA-M Series
Installation Report
Hanyang University
DVIA-M1000
DVIA-MB1000
141211R5
SEM
관련 케이스 스터디

DVIA-M Series
FEI NOVA450 DVIA-MB1000 (MB1000 P8) installation worksheet — Seoul Hanyang University — 2020-07-14
07-14-2020자세히

DVIA-M Series
Hanyang University FEI SEM DVIA-MB1000 (MB1000 P8) vibration environment measurement — 2014-12-29
12-29-2014자세히

DVIA-M Series
Chinese Academy of Sciences — Institute of Metal Research ZEISS GeminiSEM360 DVIA-M1000 (230328R2) Installation Report
05-15-2026자세히